FDA has 10 inspections on file for Preservation Solutions, Inc. in Elkhorn, Wisconsin (FEI 1000221028), from 2009 to 2025. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 3 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two of the seven FOIA requests for this site have asked FDA for this inspection's 483.

FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Nine regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
10, 2009–2025
5 OAI · 3 VAI · 2 NAI
Inspections with 483 observations shown
7 of 10
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
7 requests, latest FY2023
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

10 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Documentation of software validation

      Software validation activities and results for computers or automated data processing systems used as part of production have not been documented.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been adequately established.

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to a process and procedure have not been adequately established.

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been adequately established.

    5. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    Show 4 more citations
    1. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    2. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    3. Statistical techniques - Lack of or inadequate procedures

      Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.

    4. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate.

  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Personnel

      Personnel do not have the necessary training to perform their jobs.

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    3. Process changes - review, evaluation and revalidation

      A validated process was not reviewed and evaluated when changes or process deviations occurred.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    5. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    Show 2 more citations
    1. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures for monitoring and control of process parameters for a validated process have not been adequately established.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to a specification have not been adequately established.

    3. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

    4. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    5. Contamination control, Lack of or inadequate procedures

      Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been established.

    Show 2 more citations
    1. Evaluation and Selection, Suppliers, Contractors, etc.

      Potential suppliers were not evaluated based on their ability to meet specified requirements.

    2. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

  6. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices13 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been adequately established.

    2. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    3. Evaluation and Selection, Suppliers, Contractors, etc.

      Potential suppliers, contractors and consultants were not evaluated and selected based on their ability to meet specified requirements.

    4. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    5. Lack of System for Determining MDR Events

      The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.

    Show 8 more citations
    1. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    2. Implementing Personnel Procedures, Health, Cleanliness.

      Requirements addressing the personal practices of employees have not been implemented.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    4. Equipment Installation, Placement, Specified Requirements

      The equipment used in the manufacturing process does not meet specified requirements.

    5. Packaging

      Device packaging and/or shipping containers are not designed and constructed to protect the device from alteration or damage during processing, storage, handling, and distribution.

    6. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    7. Lack of or inadequate procedures for storage.

      Procedures for the control of storage areas and stock rooms have not been adequately established.

    8. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  7. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design output - documentation

      Design output was not documented before release.

    2. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been established.

    3. Acceptance status

      The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria .

    4. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    5. Process changes - review, evaluation and revalidation

      A validated process was not revalidated when changes or process deviations occurred.

    Show 3 more citations
    1. Evaluation and Selection, Suppliers, Contractors, etc.

      Potential suppliers and contractors were not evaluated and selected based on their ability to meet specified requirements.

    2. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    3. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been adequately established.

  8. NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
  9. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  10. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Seven requests to FDA under the Freedom of Information Act named this site, most recently on FY2023.

  • Inspection ending Jun 6, 2025OAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 11, 2025. FDAgent holds no copy that FDA posted.

    1. Sep 22, 2025FDA-FOIA-2025-11147Form 483Records released
    2. Aug 5, 2025FDA-FOIA-2025-9488Form 483· 1 of 5 sitesRecords released
  • Inspection ending Jul 14, 2023OAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Jul 17, 2024FDA-FOIA-2024-6298Form 483· 1 of 5 sitesClosed
  • Inspection ending Sep 30, 2021OAI

    2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. FY2023FDA-FOIA-2023-799EIRClosed
    2. FY2023FDA-FOIA-2023-5997Form 483Closed
  • Inspection ending Mar 12, 2019OAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Aug 12, 2026. FDAgent holds no copy that FDA posted.

    1. FY2023FDA-FOIA-2023-811EIRClosed
    2. Aug 3, 2026FDA-FOIA-2026-7460Form 483Records released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 14 sections of 21 CFR in FDA’s published citations for 7 of this facility’s 10 inspections, with how many of them cited each.

Show all 14

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000221028 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000221028. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.