Skip to content

Zimmer, Inc.

1800 W Center St, Warsaw, Indiana, United States

FEI 1000220733

FDA has 10 inspections on file for Zimmer, Inc. in Warsaw, Indiana (FEI 1000220733), from 2009 to 2022. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

The most recent OAI inspection ended on Nov 20, 2015. 1 import refusal is filed under this FEI, on . No warning letter or court action is filed under this FEI.

12 regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
10, 2009–2022
3 OAI · 4 VAI · 3 NAI
Inspections with 483 observations shown
7 of 10
2 posted 483s not matched to an inspection
Form 483s posted by FDA
↗
2 posted
Warning letters
None on file
Import refusals
1 refusal
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

10 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  2. Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated have not been adequately established.

    2. Lack of Written MDR Procedures

      Written MDR procedures have not been implemented.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Software validation for automated processes

      Software used as part of the quality system has not been adequately validated for its intended use according to an established protocol.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    Show 3 more citations
    1. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    2. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures for monitoring and control of process parameters for a validated process have not been adequately established.

    3. Packaging

      Device packaging and/or shipping containers are not designed and constructed to protect the device from alteration or damage during processing, storage, handling, and distribution.

  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been adequately established.

    2. Lack of System for Determining MDR Events

      The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.

    3. Packaging

      Device packaging and/or shipping containers are not designed and constructed to protect the device from alteration or damage during processing, storage, handling, and distribution.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    5. Sampling plans

      Sampling plans are not based on valid statistical rationale .

    Show 5 more citations
    1. Remedial action

      When test/measurement equipment was found to not meet accuracy and precision limits, inadequate action was taken to evaluate whether there was any adverse effect on the device's quality .

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

    3. Design output - documentation

      Design output was not documented before release.

    4. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    5. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

  5. Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
  6. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality System Procedures

      Quality system procedures and instructions have not been established.

    2. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    3. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    5. Design history file

      The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.

    Show 7 more citations
    1. Records of complaint investigation

      Records of complaint investigations do not include required information.

    2. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated have not been established.

    3. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    4. Packaging

      Device packaging and/or shipping containers are not designed and constructed to protect the device from alteration or damage during processing, storage, handling, and distribution.

    5. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been established to control product that does not conform to specified requirements.

    6. Sampling plans

      Sampling plans are not based on valid statistical rationale.

    7. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been established.

  7. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Bioresearch Monitoring2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Sponsor securing investigator compliance

      An investigator was not complying with the signed agreement and investigational plan and compliance of the investigator was not promptly secured.

    2. Reasonably foreseeable risks or discomforts

      The informed consent document lacked a description of reasonably foreseeable risks or discomforts to the subject.

  8. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
  9. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been established.

    2. Design output - Lack of or inadequate procedures

      Procedures for design output have not been adequately established.

    3. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated, inspected, checked, and maintained have not been adequately established.

    4. Packaging

      Device packaging and/or shipping containers are not designed and constructed to protect the device from alteration or damage during processing, storage, handling, and distribution.

    5. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    Show 5 more citations
    1. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    3. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    5. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures for monitoring and control of process parameters for a validated process have not been adequately established.

  10. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    2. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    3. DMR - not or inadequately maintained

      A device master record has not been maintained.

    4. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been established to control product that does not conform to specified requirements.

    5. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

  11. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation - user needs and intended uses

      Design validation did not ensure the device conforms to defined user needs and intended uses.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    3. Management ensuring quality policy is understood

      Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.

    4. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with the device master record.

    5. Lack of or inadequate In-process acceptance procedures

      Procedures for the acceptance of in-process product have not been established.

    Show 5 more citations
    1. Design input - Lack of or inadequate procedures

      Procedures for design input have not been established.

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

    3. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been established.

    4. Design output - Lack of or inadequate procedures

      Procedures for design output have not been established.

    5. Equipment suitability & capability

      Certain inspection, measuring, and test equipment is not suitable for its intended purposes.

  12. NAINo objectionable conditions that need follow-upDevices · Project Evaluation: DevicesNone

Regulations FDA cited here

The 19 sections of 21 CFR in FDA’s published citations for 7 of this facility’s 10 inspections, with how many of them cited each.

Show all 19

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1000220733, one row per refused product line.

1 refusal on 2018-10-23 · charged NOT LISTED on 1, FRNMFGREG on 1 · no FDA sample analysis

FDA firm profile
  • 2018-10-23

    ARTHROSCOPE AND ACCESSORIES87HRX

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast ImportsAEK-5047811-1/11/1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000220733 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000220733. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.