Zimmer Inc. Form 483, July 2011

FDA issued this Form 483 to Zimmer Inc. in Warsaw, Indiana (FEI 1000220733) at the end of a device inspection that ended July 22, 2011. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Jan 17, 2024. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Jul 22, 2011
Devices · Compliance: Devices
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Jan 17, 2024
4,562 days after the inspection
Facility
FEI 1000220733
Warsaw, Indiana

FDA's 10 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. Design validation - user needs and intended uses

    Design validation did not ensure the device conforms to defined user needs and intended uses.

  2. Lack of or inadequate procedures

    Procedures for corrective and preventive action have not been established.

  3. Management ensuring quality policy is understood

    Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.

  4. DHR content

    The device history record does not demonstrate that the device was manufactured in accordance with the device master record.

  5. Lack of or inadequate In-process acceptance procedures

    Procedures for the acceptance of in-process product have not been established.

  6. Design input - Lack of or inadequate procedures

    Procedures for design input have not been established.

  7. Lack of or inadequate process validation

    A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

  8. Lack of or inadequate receiving acceptance procedures

    Procedures for acceptance of incoming product have not been established.

  9. Design output - Lack of or inadequate procedures

    Procedures for design output have not been established.

  10. Equipment suitability & capability

    Certain inspection, measuring, and test equipment is not suitable for its intended purposes.

Other Form 483s FDA posted for this facility

Every Form 483 FDA posts under FEI 1000220733, newest first.

All inspections at this facility →

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.