Zimmer Biomet, Inc. Form 483, October 2017
FDA issued this Form 483 to Zimmer Biomet, Inc. in Warsaw, Indiana (FEI 1000220733) at the end of a device inspection that ended October 16, 2017. FDA classified the inspection Voluntary Action Indicated (VAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Jan 17, 2024. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Oct 16, 2017
- Devices · Postmarket Assurance: Devices
- FDA's outcome
- VAI
- Voluntary Action Indicated
- Posted by FDA
- Jan 17, 2024
- 2,284 days after the inspection
FDA's 8 citations from this inspection
FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Lack of Written MDR Procedures
Written MDR procedures have not been implemented.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Software validation for automated processes
Software used as part of the quality system has not been adequately validated for its intended use according to an established protocol.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures for monitoring and control of process parameters for a validated process have not been adequately established.
Packaging
Device packaging and/or shipping containers are not designed and constructed to protect the device from alteration or damage during processing, storage, handling, and distribution.
Other Form 483s FDA posted for this facility
Every Form 483 FDA posts under FEI 1000220733, newest first.
- Form 483, July 2011Inspection ended Jul 22, 2011 · Posted Jan 17, 2024VAI
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.