MIRA, Inc.
414 Quaker Hwy, Uxbridge, Massachusetts, United States
FEI 1218813
FDA has 10 inspections on file for MIRA, Inc. in Uxbridge, Massachusetts (FEI 1218813), from 2009 to 2025. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
No warning letter, court action or import refusal is filed under this FEI.
17 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 10, 2009–2025
- 0 OAI · 8 VAI · 2 NAI
- Inspections with 483 observations shown
- 8 of 10
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 4 requests, latest Apr 15, 2025
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
10 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated, inspected, checked and maintained have not been adequately established.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Document change records, maintained.
Records of changes to documents were not adequately maintained.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design history file
The design history file does not demonstrate that the design was developed following the approved design plan.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Contamination control, Lack of or inadequate procedures
Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been adequately established.
Document change records, maintained.
Records of changes to documents were not maintained.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
Design validation - simulated testing
The design was not validated under actual or simulated use conditions.
Statistical techniques - Lack of or inadequate procedures
Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been established.
Show 5 more citations
Quality audits - auditor independence
Individuals who conduct quality audits have direct responsibility for the matters being audited.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers have not been established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Management review - Lack of or inadequate procedures
Procedures for management review have not been adequately established.
Show 6 more citations
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers have not been adequately established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Design history file
The design history file was not established.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Training records
Personnel training is not documented.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Show 5 more citations
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Availability
Required records are not stored to minimize deterioration and prevent loss.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices23 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Environmental control system-documentation/review inspection
Inspections of environmental control system were not reviewed.
Incoming inspection, testing, verification
Incoming product was not adequately inspected or tested to verify conformance to specifications.
Nonconforming product, Lack of or inadequate procedures
Procedures were not followed for the control of products that do not conform to specifications.
Training records
Employee training is not fully documented.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Show 18 more citations
Documentation
Process validation activities and results have not been fully documented.
Documentation - review in response to changes or deviations
There is no documentation of the revalidation of a process conducted in response to changes or process deviations.
Documentation of software validation
Software validation activities for computers or automated data processing systems used as part of production have not been fully documented.
Test and inspection data
Service reports do not include applicable test and/or inspection data.
Documentation of validated process performance
There is no documentation of the major equipment used for a validated process.
Management representative
No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.
Time to report - 10 days
A report of the required information regarding device correction and removal actions was not sent to FDA within 10 days of initiating the correction or removal.
CAPA verification/validation of corrective/preventive action
Corrective and preventive actions have not been verified or validated to ensure that the action is effective and does not adversely affect the finished device.
Lack of or inadequate instructions
Installation, inspection, and test procedures have not been established, documented, and implemented such that the device will perform as intended after installation.
Lack of procedures, or not maintained
Document control procedures were not implemented.
Lack of or inadequate organizational structure
The organization structure is not adequate to assure that quality system requirements are fully met.
Production processes
Production processes were not complete and controlled to ensure that a device conforms to its specifications.
Design control - no procedures
Procedures to control the design process of the device were not defined, documented, complete, and implemented.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not implemented.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
DHR content
The device history record does not demonstrate the device is manufactured in accordance with the device master record.
DMR QA procedures and specifications
The device master record does not include or refer to the location of all quality assurance procedures and specifications.
Environmental control Lack of or inadequate procedures
Appropriate procedures have not been defined for controlling environmental conditions.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Analyzing service report
Service reports were not analyzed following appropriate statistical methods.
Servicing - Lack of or inadequate procedures
Servicing procedures have not been established.
Quality audits - auditor independence
Individuals who conduct quality audits have direct responsibility for the matters being audited.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Four requests to FDA under the Freedom of Information Act named this site, most recently on Apr 15, 2025. Each one asked for a Form 483.
Inspection ending Jun 12, 2024
2 requests. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.
- Sep 4, 2024FDA-FOIA-2024-7838Form 483Closed
- Aug 2, 2024FDA-FOIA-2024-6891Form 483· 1 of 3 sitesClosed
Inspection ending Nov 18, 2021NAI
2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Apr 15, 2025FDA-FOIA-2025-3411Form 483Closed
- Dec 22, 2023FDA-FOIA-2023-11425Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 21 sections of 21 CFR in FDA’s published citations for 8 of this facility’s 10 inspections, with how many of them cited each.
- 21 CFR 820.70Production and process controls7 of 8
- 21 CFR 820.75Process validation5 of 8
- 21 CFR 820.100Corrective and preventive action5 of 8
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?4 of 8
- 21 CFR 820.30Design controls4 of 8
- 21 CFR 820.50Purchasing controls4 of 8
- 21 CFR 820.22Quality audit3 of 8
- 21 CFR 820.25Personnel3 of 8
Show all 21
- 21 CFR 820.40[Reserved]3 of 8
- 21 CFR 820.20Management responsibility2 of 8
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 8
- 21 CFR 820.90Nonconforming product2 of 8
- 21 CFR 820.181Device master record2 of 8
- 21 CFR 820.184Device history record2 of 8
- 21 CFR 820.198Complaint files2 of 8
- 21 CFR 820.200Servicing2 of 8
- 21 CFR 806.10Reports of corrections and removals1 of 8
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 8
- 21 CFR 820.170Installation1 of 8
- 21 CFR 820.180General requirements1 of 8
- 21 CFR 820.250Statistical techniques1 of 8
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1218813 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1218813. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.