Transfer Technology, Inc.
5702 Bolsa Ave Ste 200, Huntington Beach, California, United States
FEI 3009494532DUNS 037968132
FDA has 5 inspections on file for Transfer Technology, Inc. in Huntington Beach, California (FEI 3009494532), from 2013 to 2025. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 3 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One of the three FOIA requests for this site has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Seven regulations were cited at more than one inspection; they are listed below.
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, packing, labeling, relabeling and repacking, through Dec 31, 2026. Its registrant is Transfer Technology (D-U-N-S 037968132).
Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 5, 2013–2025
- 0 OAI · 3 VAI · 2 NAI
- Inspections with 483 observations shown
- 5 of 5
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest FY2023
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Sampling plans
Sampling plans are not based on valid statistical rationale.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedures not in writing, fully followed
The responsibilities and procedures applicable to the quality control unit are not fully followed.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures for labeling
Procedures to control labeling activities have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design control - no procedures
Procedures for design control have not been established.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Sampling plans
Sampling plans are not written and based on valid statistical rationale.
Show 2 more citations
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Documentation
In-process inspections, tests, or other verification activities and approvals were not documented.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Sampling plans
Sampling plans are not based on valid statistical rationale.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Design input - documentation
Design input requirements were not adequately documented.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Show 7 more citations
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Documented evaluation
The evaluation of potential suppliers was not documented.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Procedures not adequately established or maintained
Document control procedures have not been adequately maintained.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been adequately established.
Design validation - production units
The design was not validated using initial production units, lots or batches or their equivalents.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Complainant data
Records of complaint investigations do not include the name, address, and phone number of the complainant.
Design transfer - Lack of or inadequate procedures
Procedures were not established to ensure that the device design was correctly transferred into production.
DMR production process specifications
The device master record does not include or refer to the location of all production and process specifications.
Design validation- Lack of or inadequate procedures
Procedures for validating the device design were not established.
Design validation - production units
The design was not validated using production units or their equivalents.
Show 7 more citations
Production monitoring and control
During production, process parameters and component and device characteristics are not fully monitored and controlled.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure that equipment is routinely calibrated, inspected, checked, and maintained were not established.
General
Complaint handling procedures for receiving, reviewing, and evaluating complaints have not been defined and completed.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Design plans - Lack of or inadequate
The design plan does not describe the design and development activities, define responsibility for implementation of design and development activities, and identify and describe the interfaces with other groups or activities as appropriate.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on FY2023. Each one asked for a Form 483.
Inspection ending Apr 30, 2025NAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 5, 2025. FDAgent holds no copy that FDA posted.
- Jul 9, 2025FDA-FOIA-2025-8457Form 483· 1 of 5 sitesRecords released
Inspection ending Apr 28, 2023VAI
2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- FY2023FDA-FOIA-2023-7703Form 483Closed
- Aug 1, 2024FDA-FOIA-2024-6697Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 17 sections of 21 CFR in FDA’s published citations for this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls4 of 5
- 21 CFR 820.181Device master record3 of 5
- 21 CFR 820.250Statistical techniques3 of 5
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 5
- 21 CFR 820.100Corrective and preventive action2 of 5
- 21 CFR 820.184Device history record2 of 5
- 21 CFR 211.22Responsibilities of quality control unit1 of 5
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 5
Show all 17
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 5
- 21 CFR 820.40[Reserved]1 of 5
- 21 CFR 820.50Purchasing controls1 of 5
- 21 CFR 820.70Production and process controls1 of 5
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 5
- 21 CFR 820.75Process validation1 of 5
- 21 CFR 820.90Nonconforming product1 of 5
- 21 CFR 820.120Device labeling1 of 5
- 21 CFR 820.198Complaint files1 of 5
FDA drug establishment registration
FDA registration search- Registered operations
- Manufacturing, packing, labeling, relabeling, repacking
- Registration runs through
- Dec 31, 2026
- Registrant
- Transfer TechnologyD-U-N-S 037968132
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3009494532 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3009494532. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.