Trilliant Surgical, LLC
727 N Shepherd Dr Ste 100, Houston, Texas, United States
FEI 3007420745
FDA has 6 inspections on file for Trilliant Surgical, LLC in Houston, Texas (FEI 3007420745), from 2010 to 2019. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 5 citations from this inspection: 4 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
The most recent OAI inspection ended on Feb 12, 2015. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Six regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 6, 2010–2019
- 2 OAI · 3 VAI · 1 NAI
- Inspections with 483 observations shown
- 6 of 6
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Sep 18, 2026
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigation of device failures
Complaints involving the possible failure of a device, labeling and packaging to meet any of its specifications were not reviewed, evaluated and investigated where necessary.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been adequately established.
Documentation
Process validation activities have not been adequately documented.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Evaluation Codes
An individual medical device manufacturer report submitted per FDA Form 3500A did not include in Block H the evaluation codes (including event codes, method of evaluation, result and conclusion codes).
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate In-process acceptance procedures
Procedures for the acceptance and control of in-process product have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Design output - review and approval
Design output was not reviewed and approved before release.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Show 5 more citations
Supplier oversight
The type and extent of control to be exercised over suppliers, contractors and consultants was not clearly defined.
Not approved or obsolete document retrieval
Documents that were not approved were observed at a location where they are being used.
Equipment control activity documentation
Equipment calibrations, inspections and checks have not been documented.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Statistical techniques - Lack of or inadequate procedures
Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Documentation
Process validation activities have not been adequately documented.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
Change records, content
Records of changes did not include a description of the change.
Show 4 more citations
Statistical techniques - Lack of or inadequate procedures
Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Documented evaluation
The evaluation of potential contractors was not documented.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Records, DHR {see also 820.184(e)}
Labels and labeling used for each finished product, lot, or batch, were not sufficiently documented in the DHR.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality audits - auditor independence
Individuals who conduct quality audits have direct responsibility for the matters being audited.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities have not been documented, including investigations of causes of nonconformities and the actions needed to correct or prevent recurrence of nonconforming product and other quality problems.
Approval, inadequate purchasing data
Purchasing information that clearly includes/describes the specified requirements for purchased or otherwise received products and services was not established, defined, and documented.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
CAPA verification/validation of corrective/preventive action
Corrective and preventive actions have not been verified or validated to ensure that the action is effective and does not adversely affect the finished device.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Sep 18, 2026. It asked for a Form 483.
Inspection ending Dec 11, 2019VAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Oct 9, 2026. FDAgent holds no copy that FDA posted.
- Sep 18, 2026FDA-FOIA-2026-9074Form 483Records released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 16 sections of 21 CFR in FDA’s published citations for this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls4 of 6
- 21 CFR 820.50Purchasing controls4 of 6
- 21 CFR 820.75Process validation3 of 6
- 21 CFR 820.100Corrective and preventive action3 of 6
- 21 CFR 820.40[Reserved]2 of 6
- 21 CFR 820.70Production and process controls2 of 6
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 6
- 21 CFR 820.198Complaint files2 of 6
Show all 16
- 21 CFR 820.250Statistical techniques2 of 6
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 6
- 21 CFR 803.52If I am a manufacturer, what information must I submit in my individual adverse event reports?1 of 6
- 21 CFR 806.10Reports of corrections and removals1 of 6
- 21 CFR 820.22Quality audit1 of 6
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 6
- 21 CFR 820.90Nonconforming product1 of 6
- 21 CFR 820.120Device labeling1 of 6
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3007420745 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3007420745. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.