Artivion, Inc
1655 Roberts Blvd Nw, Kennesaw, Georgia, United States
FEI 3001451326
FDA has 14 inspections on file for Artivion, Inc in Kennesaw, Georgia (FEI 3001451326), from 2008 to 2026. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Mar 19, 2014. FDA issued a warning letter on (CDRH). 1 import refusal is filed under this FEI, on . No court action is filed under this FEI.
Six regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 14, 2008–2026
- 2 OAI · 2 VAI · 10 NAI
- Inspections with 483 observations shown
- 4 of 14
- 1 posted 483 not matched to an inspection
- Form 483s posted by FDA
- ↗
- Warning letters
- · CDRH ↗
- Import refusals
- 1 refusal
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
14 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- NAINo objectionable conditions that need follow-upDevices · Bioresearch MonitoringNone
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation - simulated testing
The design was not validated under actual or simulated use conditions.
Equipment design and installation
Equipment used in the manufacturing process has not been appropriately designed, constructed, placed and installed to facilitate maintenance, adjustment, cleaning, and use.
Equipment control activity documentation
Equipment calibrations, inspections, checks and maintenance activities have not been documented.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely checked and maintained have not been adequately established.
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies17 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Contamination, mix ups, improper release
Storage areas and stock rooms were not controlled to prevent mix-ups, contamination and cross contamination of HCT/Ps, supplies and reagents and to prevent HCT/Ps from improperly being made available for distribution.
Lack of or inadequate procedures for handling
Procedures for product handling have not been adequately established.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not evaluated where necessary.
Statistical techniques - Lack of or inadequate procedures
Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to a method have not been adequately established.
Show 12 more citations
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been established.
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.
Lack of or inadequate procedures for storage.
Procedures for the control of storage areas and stock rooms have not been adequately established.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Documentation
Process validation activities and results have not been adequately documented.
Personnel
Personnel do not have the necessary training to perform their jobs.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of procedures, or not maintained
Document control procedures have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Equipment suitability & capability
Certain inspection and test equipment is not suitable for its intended purposes.
FDA not notified of defect or noncompliance
You did not notify the Secretary, in accordance with 21 CFR 1003.20, of an electronic product which was defective which you produced and distributed.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Show 5 more citations
Lack of or inadequate In-process acceptance procedures
Procedures for the acceptance and control of in-process product have not been adequately established.
Reproductions of warnings
Adequate instructions for service adjustments and service procedures were not provided or caused to be provided for each laser model, including legible reproductions of required labels and hazard warnings.
50 mm classification warning
The following warning was not included in the user information: "CAUTION --- The use of optical instruments with this product will increase eye hazard", when use of a 50 millimeter diameter aperture stop in a laser product classified with a 7 millimeter diameter aperture stop resulted in a higher classification of the product.
Failure to submit
You did not submit an annual report by the September 1 deadline for products requiring one.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
- OAISignificant objectionable conditions; FDA recommended regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design history file
The design history file does not demonstrate that the design was developed following the approved design plan and does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Documentation
Process validation activities and results have not been adequately documented.
Documentation of validated process performance
There is inadequate documentation of monitoring and control methods and data for a validated process.
Sampling plans
Sampling plans are not based on valid statistical rationale.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Show 8 more citations
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Production processes
Production processes were not controlled and monitored to ensure that a device conforms to its specifications.
Process changes - review, evaluation and revalidation
A validated process was not reviewed and evaluated and revalidated when changes or process deviations occurred.
Design review - documentation
The design review results, including identification of the design and the individual(s) performing the review, were not documented in the design history file.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system with sufficient frequency.
Causing contamination, increasing risks
HCT/Ps were not processed in a way that does not cause contamination or cross contamination during processing.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures for monitoring and control of process parameters for a validated process have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Bioresearch MonitoringNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- NAINo objectionable conditions that need follow-upDevices · Bioresearch MonitoringNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Regulations FDA cited here
The 22 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 14 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls3 of 4
- 21 CFR 820.70Production and process controls3 of 4
- 21 CFR 820.198Complaint files3 of 4
- 21 CFR 820.72Inspection, measuring, and test equipment2 of 4
- 21 CFR 820.75Process validation2 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 4
- 21 CFR 820.100Corrective and preventive action2 of 4
- 21 CFR 820.250Statistical techniques2 of 4
Show all 22
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 4
- 21 CFR 806.10Reports of corrections and removals1 of 4
- 21 CFR 820.20Management responsibility1 of 4
- 21 CFR 820.22Quality audit1 of 4
- 21 CFR 820.25Personnel1 of 4
- 21 CFR 820.40[Reserved]1 of 4
- 21 CFR 820.90Nonconforming product1 of 4
- 21 CFR 820.140Handling1 of 4
- 21 CFR 820.150Storage1 of 4
- 21 CFR 1002.13Annual reports1 of 4
- 21 CFR 1003.10Discovery of defect or failure of compliance by manufacturer; notice requirements1 of 4
- 21 CFR 1040.10Laser products1 of 4
- 21 CFR 1271.220Processing and process controls1 of 4
- 21 CFR 1271.260Storage1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3001451326, one row per refused product line.
1 refusal on 2025-09-04 · charged NOT LISTED on 1, NO PMA on 1 · no FDA sample analysis
FDA firm profile- 2025-09-04
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT74MIH
NOT LISTED 502(o), 801(a)(3); MISBRANDING
NO PMA 501(f)(1)(B), 801(a)(3); ADULTERATION
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-1147132-6/10001/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3001451326 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3001451326. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.