Light Age, Inc.
220 Davidson Ave Ste 216, Somerset, New Jersey, United States
FEI 2246844
FDA has 6 inspections on file for Light Age, Inc. in Somerset, New Jersey (FEI 2246844), from 2012 to 2026. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 9 citations from this inspection. They are listed below. Three of the four FOIA requests for this site have asked FDA for this inspection's 483.
The most recent OAI inspection ended on Mar 11, 2020. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
13 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 6, 2012–2026
- 1 OAI · 4 VAI · 1 NAI
- Inspections with 483 observations shown
- 6 of 6
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 4 requests, latest Sep 10, 2026
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 8.3.1
Procedure for nonconforming product not documented
Procedures to define the controls and related responsibilities and authorities for the identification, documentation, segregation, evaluation, and disposition of nonconforming product were not documented.
- ISO 13485:2016 Clause 8.3.3
Appropriate action not taken for nonconforming product after delivery or use
Appropriate action was not taken to address nonconforming product detected after delivery or use had started.
- ISO 13485:2016 Clause 4.2.4
Lack of document control
Documents required by the quality management system have not been controlled.
- ISO 13485:2016 Clause 8.2.2
Complaint handling records not maintained
Complaint handling records were not adequately maintained.
- ISO 13485:2016 Clause 7.5.1
Medical device or batch records not adequate
A record for each medical device or batch of medical devices that provides specified traceability and identifies the amount manufactured and amount approved for distribution has not been established and maintained.
Show 4 more citations
- ISO 13485:2016 Clause 8.2.4
Internal audits not conducted to ensure QMS meets requirements and is effective
Internal audits have not been adequately conducted at planned intervals to determine whether the QMS conforms to planned and documented arrangements and applicable requirements.
- ISO 13485:2016 Clause 7.6
Equipment calibration or verification not performed
Equipment calibration or verification was not performed according to documented procedures.
- ISO 13485:2016 Clause 8.5.3
Preventive action records not maintained
Records of preventive actions, including the results of any investigations and actions taken, were not adequately maintained.
- ISO 13485:2016 Clause 8.4
Records of data analysis results not maintained
Records of the results of data analyses were not maintained.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices14 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Calibration procedures - content
Calibration procedures do not include specific directions and limits for accuracy and precision.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals and with sufficient frequency.
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Show 9 more citations
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.
Complaints
Complaint files are not adequately maintained.
Remedial action
When test/measurement equipment was found to not meet accuracy and precision limits, action was taken to bring the equipment into calibration and evaluate whether there was any adverse effect on the device's quality.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices17 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Servicing - Lack of or inadequate procedures
Procedures or instructions for verifying that servicing meets specified requirements have not been established.
DMR - not or inadequately maintained
A device master record has not been maintained.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Document review, approval by designated individual
Documents were not reviewed and not approved by designated individual(s) prior to issuance .
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Show 12 more citations
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are produced in accordance with 21 CFR 820.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record and 21 CFR 820.
Acceptance status
The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals and with sufficient frequency.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Lack of or inadequate DHR procedures
Procedures for device history records have not been adequately established.
Complaints
Complaint files are not maintained.
Not approved or obsolete document retrieval
Documents that were obsolete were observed at a location where they could be used.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedures not adequately established or maintained
Document control procedures have not been adequately established and maintained.
Complaints
Complaint files are not maintained.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Show 1 more citation
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system.
- NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety Act15 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Acceptance activities were not adequately documented and adequately maintained as part of the device history record.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Procedures for product review,disposition lack of/inadequate
Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.
Approval of purchasing data {see also 820.40(a)}
The purchasing data were not approved.
Show 10 more citations
Not approved or obsolete document retrieval
Documents that were not approved were observed at a location where they are being used.
Equipment control activity documentation
Equipment calibrations have not been documented.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are produced in accordance with 21 CFR 820.
Quality Audits - defined intervals
Quality audits were not performed at sufficient frequency to determine whether the quality system activities and results comply with quality system procedures.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with 21 CFR 820.
DMR - not or inadequately maintained
A device master record has not been maintained.
Personnel
Personnel do not have the necessary training to perform their jobs.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Periodic equipment inspections
Periodic inspections of equipment conducted to ensure adherence to applicable maintenance schedules were not documented.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Four requests to FDA under the Freedom of Information Act named this site, most recently on Sep 10, 2026. Each one asked for a Form 483.
Inspection ending May 15, 2026VAI
3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 2, 2026. FDAgent holds no copy that FDA posted.
- Sep 10, 2026FDA-FOIA-2026-8783Form 483Records released
- Aug 21, 2026FDA-FOIA-2026-8128Form 483· 1 of 4 sitesOpen
- Jul 29, 2026FDA-FOIA-2026-7249Form 483· 1 of 5 sitesRecords released
Inspection ending Mar 11, 2020OAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Apr 29, 2026. FDAgent holds no copy that FDA posted.
- Apr 1, 2026FDA-FOIA-2026-3201Form 483Records released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 17 sections of 21 CFR in FDA’s published citations for this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.20Management responsibility4 of 6
- 21 CFR 820.40[Reserved]4 of 6
- 21 CFR 820.72Inspection, measuring, and test equipment4 of 6
- 21 CFR 820.100Corrective and preventive action4 of 6
- 21 CFR 820.198Complaint files4 of 6
- 21 CFR 820.22Quality audit3 of 6
- 21 CFR 820.25Personnel3 of 6
- 21 CFR 820.30Design controls3 of 6
Show all 17
- 21 CFR 820.90Nonconforming product3 of 6
- 21 CFR 820.181Device master record3 of 6
- 21 CFR 820.184Device history record3 of 6
- 21 CFR 820.50Purchasing controls2 of 6
- 21 CFR 820.70Production and process controls2 of 6
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 6
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 6
- 21 CFR 820.86Acceptance status1 of 6
- 21 CFR 820.200Servicing1 of 6
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2246844 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2246844. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.