Optikem International, Inc.
11452 W Hampden Pl, Denver, Colorado, United States
FEI 1720129DUNS 092765312
FDA has 6 inspections on file for Optikem International, Inc. in Denver, Colorado (FEI 1720129), from 2010 to 2024. The latest ended on and was classified OAI, official action indicated. FDA has issued 3 warning letters under this FEI, the latest on (CDRH).
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 7 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two of the four FOIA requests for this site have asked FDA for this inspection's 483.
No court action or import refusal is filed under this FEI.
Five regulations were cited at more than one inspection; they are listed below.
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, through Dec 31, 2026. Its registrant is Optikem International, Inc. (D-U-N-S 092765312).
Sources:FDA firm profile ↗Warning letter ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 6, 2010–2024
- 2 OAI · 3 VAI · 1 NAI
- Inspections with 483 observations shown
- 6 of 6
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 4 requests, latest Oct 2, 2024
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis was not performed.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Show 2 more citations
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Design verification - Lack of or inadequate procedures
Procedures for design verification have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality Assurance14 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Valid stability test methods
The written stability program for drug products does not include reliable, meaningful and specific test methods.
Sanitation--buildings not clean, free of infestation
Buildings used in the manufacture, processing, packing or holding of drug products are not maintained in a clean and sanitary condition and free of infestation by rodents, birds insects, and other vermin.
Control procedures to monitor and validate performance
Control procedures are not established which monitor the output and validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product.
Written procedures not established/followed
Written procedures are not established and followed for the cleaning and maintenance of equipment, including utensils, used in the manufacture, processing, packing or holding of a drug product.
Procedures for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established, written and followed.
Show 9 more citations
Prepared for each batch, include complete information
Batch production and control records are not prepared for each batch of drug product produced and do not include complete information relating to the production and control of each batch.
Manufacturing Instructions and Specifications
The master production and control records are deficient in that they do not include complete manufacturing, control, instructions, sampling, testing, procedures and specifications.
Scientifically sound laboratory controls
Laboratory controls do not include the establishment of scientifically sound and appropriate specifications, standards, sampling plans and test procedures designed to assure that components and drug products conform to appropriate standards of identity, strength, quality and purity.
Validation lacking for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile did not include validation of the process.
Investigations of discrepancies, failures
There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.
Absence of Written Procedures
Your firm failed to establish adequate written procedures for production and process controls designed to assure that the drug products have the identity, strength, purity, and quality that they are purported or represented to possess.
Equipment to control conditions
Aseptic processing areas are deficient regarding systems for maintaining any equipment used to control the aseptic conditions.
Procedures not in writing, fully followed
The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed.
Equipment Design, Size and Location
Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design to facilitate operations for its intended use and cleaning and maintenance.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Labeling inspection - UDI
Labeling was not sufficiently examined by a designated individual for accuracy including unique device identifier (UDI) or universal product code (UPC).
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lab controls established, including changes
The establishment of specifications, sampling plans and test procedures including any changes thereto, are not drafted by the appropriate organizational unit and reviewed and approved by the quality control unit.
Prepared for each batch, include complete information
Batch production and control records are not prepared for each batch of drug product produced and do not include complete information relating to the production and control of each batch.
Written records kept in individual logs
Written records of major equipment cleaning, maintenance and use are not included in individual equipment logs.
Cleaning / Sanitizing / Maintenance
Equipment and utensils are not cleaned and sanitized at appropriate intervals to prevent contamination that would alter the safety, identity, strength, quality or purity of the drug product.
Sanitation--buildings not clean, free of infestation
Buildings used in the manufacture, processing, packing or holding of drug products are not maintained in a clean and sanitary condition and free of infestation by rodents, birds insects, and other vermin.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated, inspected and checked have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities have not been documented, including investigations of causes of nonconformities and implementation of corrective and preventive actions.
Calibration procedures - content
Calibration procedures do not include specific directions and limits for accuracy and precision and provisions for remedial action.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures were not implemented for monitoring and control of process parameters for validated processes.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been fully validated and approved according to established procedures.
Environmental control Lack of or inadequate procedures
Appropriate procedures have not been defined, documented, and followed for controlling environmental conditions.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Four requests to FDA under the Freedom of Information Act named this site, most recently on Oct 2, 2024. Each one asked for a Form 483.
Inspection ending Mar 1, 2024OAI
2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Oct 2, 2024FDA-FOIA-2024-8659Form 483Closed
- Jun 7, 2024FDA-FOIA-2024-5009Form 483· 1 of 3 sitesClosed
Inspection ending Jan 26, 2024OAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- May 10, 2024FDA-FOIA-2024-4290Form 483· 1 of 2 sitesClosed
Inspection ending Jul 11, 2022NAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Nov 15, 2023FDA-FOIA-2023-10236Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 22 sections of 21 CFR in FDA’s published citations for this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.72Inspection, measuring, and test equipment4 of 6
- 21 CFR 820.75Process validation3 of 6
- 21 CFR 211.56Sanitation2 of 6
- 21 CFR 211.67Equipment cleaning and maintenance2 of 6
- 21 CFR 211.160General requirements2 of 6
- 21 CFR 211.188Batch production and control records2 of 6
- 21 CFR 820.70Production and process controls2 of 6
- 21 CFR 820.100Corrective and preventive action2 of 6
Show all 22
- 21 CFR 211.22Responsibilities of quality control unit1 of 6
- 21 CFR 211.42Design and construction features1 of 6
- 21 CFR 211.63Equipment design, size, and location1 of 6
- 21 CFR 211.100Written procedures; deviations1 of 6
- 21 CFR 211.110Sampling and testing of in-process materials and drug products1 of 6
- 21 CFR 211.113Control of microbiological contamination1 of 6
- 21 CFR 211.166Stability testing1 of 6
- 21 CFR 211.182Equipment cleaning and use log1 of 6
- 21 CFR 211.186Master production and control records1 of 6
- 21 CFR 211.192Production record review1 of 6
- 21 CFR 820.30Design controls1 of 6
- 21 CFR 820.120Device labeling1 of 6
- 21 CFR 820.184Device history record1 of 6
- 21 CFR 820.198Complaint files1 of 6
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
- Warning letterFormal notice of significant violationsCDER · Division of Pharmaceutical Quality Operations IV · DrugsLetter on fda.govFDA record
- Warning letterFormal notice of significant violationsCDER · DrugsLetter on fda.govFDA record
FDA drug establishment registration
FDA registration search- Registered operations
- Manufacturing
- Registration runs through
- Dec 31, 2026
- Registrant
- Optikem International, Inc.D-U-N-S 092765312
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1720129 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1720129. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.