Invacare Corporation
1200 Taylor St, Elyria, Ohio, United States
FEI 1525712
FDA has 9 inspections on file for Invacare Corporation in Elyria, Ohio (FEI 1525712), from 2008 to 2023. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Jun 24, 2021. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Seven regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 9, 2008–2023
- 4 OAI · 2 VAI · 3 NAI
- Inspections with 483 observations shown
- 3 of 9
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Dec 7, 2023
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
9 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation - Risk analysis
Results of the design risk analysis were not adequately documented.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Documentation of validated process performance
There is no documentation of monitoring and control methods and data and the individual performing the process for a validated process.
Documentation
Process validation activities and results have not been adequately documented and adequately approved.
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.
Show 4 more citations
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Lack of or inadequate In-process acceptance procedures
Procedures for the acceptance of in-process product have not been adequately established.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices14 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Production processes
Production processes were not developed, conducted, controlled, and monitored to ensure that a device conforms to its specifications.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Show 9 more citations
Lack of or inadequate procedures for distribution
Procedures for control and distribution of finished devices have not been adequately established.
Report of death, injury
The importer failed to submit a report to FDA on FDA form 3500A, with a copy to the manufacturer, within 30 days after receiving information that one of its marketed devices may have caused or contributed to a serious injury.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Acceptance status
The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria.
Software validation for automated processes
Software used as part of production has not been validated for its intended use according to an established protocol.
Documentation
Process validation have not been .
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Time to report - 10 days
A report of the required information regarding device correction and removal actions was not sent to FDA within 10 days of initiating the correction or removal.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated and approved according to established procedures.
Report of death, injury
The importer failed to submit a report to FDA on FDA form 3500A, with a copy to the manufacturer, within 30 days after receiving information that one of its marketed devices may have caused or contributed to a serious injury.
Design verification - Lack of or inadequate procedures
Design verification did not confirm that the design output meets the design input requirements.
Show 6 more citations
Design control - no procedures
Procedures to control the design process of the device were not implemented.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Lack of System for Timely Submission of Reports
The written MDR procedure does not include an internal system which provides for timely transmission of complete medical device reports to.
Quality Audit/Reaudit - conducted
Quality audits were not conducted at sufficient regular intervals, as prescribed by internal procedures to verify that the quality system is effective in fulfilling your quality system objectives.
Documentation
Corrective and preventive action activities have not been documented, including implementation of corrective and preventive actions.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures were not established for monitoring and control of process parameters for validated processes.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Dec 7, 2023. It asked for a Form 483.
Inspection ending Jan 31, 2020NAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Dec 7, 2023FDA-FOIA-2023-10930Form 483No records found
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 14 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 9 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls3 of 3
- 21 CFR 820.75Process validation3 of 3
- 21 CFR 820.100Corrective and preventive action3 of 3
- 21 CFR 803.40If I am an importer, what reporting requirements apply to me?2 of 3
- 21 CFR 820.70Production and process controls2 of 3
- 21 CFR 820.90Nonconforming product2 of 3
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 3
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 3
Show all 14
- 21 CFR 806.10Reports of corrections and removals1 of 3
- 21 CFR 820.22Quality audit1 of 3
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 3
- 21 CFR 820.86Acceptance status1 of 3
- 21 CFR 820.160Distribution1 of 3
- 21 CFR 820.198Complaint files1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
Other FEIs recorded as Invacare Corporation
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Invacare CorporationFEI 3002889431Elyria, Ohio, United States
Last inspected 2023-03-30NAI
Invacare CorporationFEI 1000220413North Ridgeville, Ohio, United States
Last inspected 2023-03-30NAI
Invacare CorporationFEI 1031452Sanford, Florida, United States
Last inspected 2017-11-27NAI
Invacare CorporationFEI 1000132244Pinellas Park, Florida, United States
Last inspected 2017-11-07VAI
Invacare CorporationFEI 3005031582Elyria, Ohio, United States
Last inspected 2010-12-17OAI
Invacare CorporationFEI 1219590Elyria, Ohio, United States
Last inspected 2010-12-06NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1525712 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1525712. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.