Invacare Corporation

1 Invacare Way, Elyria, Ohio, United States

FEI 3002889431

FDA has 10 inspections on file for Invacare Corporation in Elyria, Ohio (FEI 3002889431), from 2008 to 2023. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two of the three FOIA requests for this site asked FDA for the 483 from a later inspection, ending Apr 9, 2025, that FDA's Data Dashboard doesn't list yet.

The most recent OAI inspection ended on Jun 24, 2021. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

14 regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
10, 2008–2023
4 OAI · 3 VAI · 3 NAI
Inspections with 483 observations shown
7 of 10
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
3 requests, latest FY2023
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

10 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate and is incomplete.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    2. Statistical techniques - Lack of or inadequate procedures

      Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.

    3. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    4. Procedures not adequately established or maintained

      Document control procedures have not been adequately established.

    5. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    Show 1 more citation
    1. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    2. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    3. Document change records, maintained.

      Records of changes to documents were not adequately maintained.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
  5. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  6. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

  7. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design input - documentation

      Design input requirements were not adequately documented.

    2. n/a

      Procedural requirements

      Deviation from the procedural requirements of a decree of injunction.

    3. Design verification - output does not meet input requirement

      Design verification does not confirm that design output meets design input requirements.

    4. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    5. Documentation of validated process performance

      There is inadequate documentation of monitoring and control methods and data for a validated process.

    Show 3 more citations
    1. Design validation - user needs and intended uses

      Design validation did not ensure the device conforms to defined user needs and intended uses.

    2. Documentation

      Process validation activities have not been adequately documented.

    3. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate.

  8. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design output - Lack of or inadequate procedures

      Procedures for design output have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    4. Complaints

      Complaint files are not adequately maintained.

    5. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    Show 7 more citations
    1. Personnel

      Personnel do not have the necessary training to perform their jobs.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Lack of System for Event Evaluations

      The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication, and evaluation of events that may be subject to medical device reporting requirements.

    4. Design input - documentation

      Design input requirements were not adequately documented.

    5. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    6. Report of malfunction likely to cause death or injury

      The importer failed to submit a report to the manufacturer on FDA Form 3500A within 30 days concerning information that one of the devices marketed by the importer has malfunctioned and that such device or a similar device marketed by the importer would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    7. Sampling methods - Lack of or inadequate procedures

      Procedures to ensure sampling methods are adequate for their intended use have not been adequately established.

  9. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices18 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of Written MDR Procedures

      Written MDR procedures have not been developed and maintained.

    2. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.

    3. Personnel

      Personnel do not have the necessary training to perform their jobs.

    4. Design input - documentation

      Design input requirements were not adequately documented.

    5. Design control - no procedures

      Procedures for design control have not been established.

    Show 13 more citations
    1. Design verification - Lack of or inadequate procedures

      Procedures for design verification have not been adequately established.

    2. Review and evaluation for investigation

      Not all complaints have been adequately reviewed and evaluated to determine whether an investigation is necessary.

    3. Design verification - output does not meet input requirement

      Design verification does not confirm that design output meets design input requirements.

    4. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    5. Design output - Lack of or inadequate procedures

      Procedures for design output have not been adequately established.

    6. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    7. Lack of System for Event Evaluations

      The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication, and evaluation of events that may be subject to medical device reporting requirements.

    8. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    9. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    10. Design validation - user needs and intended uses

      Design validation did not ensure the device conforms to defined user needs and intended uses.

    11. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been established.

    12. Documentation

      Corrective and preventive action activities and/or results have not been documented.

    13. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  10. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Three requests to FDA under the Freedom of Information Act named this site, most recently on FY2023. Each one asked for a Form 483.

  • Inspection ending Apr 9, 2025

    2 requests. Not listed in FDA’s Data Dashboard yet. FDA released records in answer to a request for its Form 483 on Jul 21, 2025. FDAgent holds no copy that FDA posted.

    1. Oct 15, 2025FDA-FOIA-2025-12028Form 483· 1 of 4 sitesRecords released
    2. May 16, 2025FDA-FOIA-2025-4483Form 483· 1 of 4 sitesRecords released
  • Inspection ending Jun 24, 2021OAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. FY2023FDA-FOIA-2023-8175Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 12 sections of 21 CFR in FDA’s published citations for 7 of this facility’s 10 inspections, with how many of them cited each.

Show all 12

Warning letters and compliance actions

FDA Data Dashboard compliance actions

Other FEIs recorded as Invacare Corporation

Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3002889431 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3002889431. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.