Invacare Corporation
1 Invacare Way, Elyria, Ohio, United States
FEI 3002889431
FDA has 10 inspections on file for Invacare Corporation in Elyria, Ohio (FEI 3002889431), from 2008 to 2023. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two of the three FOIA requests for this site asked FDA for the 483 from a later inspection, ending Apr 9, 2025, that FDA's Data Dashboard doesn't list yet.
The most recent OAI inspection ended on Jun 24, 2021. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
14 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 10, 2008–2023
- 4 OAI · 3 VAI · 3 NAI
- Inspections with 483 observations shown
- 7 of 10
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest FY2023
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
10 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate and is incomplete.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Statistical techniques - Lack of or inadequate procedures
Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Show 1 more citation
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Document change records, maintained.
Records of changes to documents were not adequately maintained.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design input - documentation
Design input requirements were not adequately documented.
- n/a
Procedural requirements
Deviation from the procedural requirements of a decree of injunction.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Documentation of validated process performance
There is inadequate documentation of monitoring and control methods and data for a validated process.
Show 3 more citations
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Documentation
Process validation activities have not been adequately documented.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design output - Lack of or inadequate procedures
Procedures for design output have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Complaints
Complaint files are not adequately maintained.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Show 7 more citations
Personnel
Personnel do not have the necessary training to perform their jobs.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication, and evaluation of events that may be subject to medical device reporting requirements.
Design input - documentation
Design input requirements were not adequately documented.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Report of malfunction likely to cause death or injury
The importer failed to submit a report to the manufacturer on FDA Form 3500A within 30 days concerning information that one of the devices marketed by the importer has malfunctioned and that such device or a similar device marketed by the importer would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Sampling methods - Lack of or inadequate procedures
Procedures to ensure sampling methods are adequate for their intended use have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices18 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of Written MDR Procedures
Written MDR procedures have not been developed and maintained.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.
Personnel
Personnel do not have the necessary training to perform their jobs.
Design input - documentation
Design input requirements were not adequately documented.
Design control - no procedures
Procedures for design control have not been established.
Show 13 more citations
Design verification - Lack of or inadequate procedures
Procedures for design verification have not been adequately established.
Review and evaluation for investigation
Not all complaints have been adequately reviewed and evaluated to determine whether an investigation is necessary.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Design output - Lack of or inadequate procedures
Procedures for design output have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication, and evaluation of events that may be subject to medical device reporting requirements.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been established.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on FY2023. Each one asked for a Form 483.
Inspection ending Apr 9, 2025
2 requests. Not listed in FDA’s Data Dashboard yet. FDA released records in answer to a request for its Form 483 on Jul 21, 2025. FDAgent holds no copy that FDA posted.
- Oct 15, 2025FDA-FOIA-2025-12028Form 483· 1 of 4 sitesRecords released
- May 16, 2025FDA-FOIA-2025-4483Form 483· 1 of 4 sitesRecords released
Inspection ending Jun 24, 2021OAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- FY2023FDA-FOIA-2023-8175Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for 7 of this facility’s 10 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls4 of 7
- 21 CFR 820.100Corrective and preventive action4 of 7
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?3 of 7
- 21 CFR 820.50Purchasing controls3 of 7
- 21 CFR 820.198Complaint files3 of 7
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 7
- 21 CFR 820.25Personnel2 of 7
- 21 CFR 820.40[Reserved]2 of 7
Show all 12
- 21 CFR 820.250Statistical techniques2 of 7
- 21 CFR 803.40If I am an importer, what reporting requirements apply to me?1 of 7
- 21 CFR 820.75Process validation1 of 7
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 7
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
Other FEIs recorded as Invacare Corporation
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Invacare CorporationFEI 1525712Elyria, Ohio, United States
Last inspected 2023-03-30NAI
Invacare CorporationFEI 1000220413North Ridgeville, Ohio, United States
Last inspected 2023-03-30NAI
Invacare CorporationFEI 1031452Sanford, Florida, United States
Last inspected 2017-11-27NAI
Invacare CorporationFEI 1000132244Pinellas Park, Florida, United States
Last inspected 2017-11-07VAI
Invacare CorporationFEI 3005031582Elyria, Ohio, United States
Last inspected 2010-12-17OAI
Invacare CorporationFEI 1219590Elyria, Ohio, United States
Last inspected 2010-12-06NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3002889431 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3002889431. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.