Invacare Corporation

2101 E Lake Mary Blvd, Sanford, Florida, United States

FEI 1031452

FDA has 5 inspections on file for Invacare Corporation in Sanford, Florida (FEI 1031452), from 2009 to 2017. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

The most recent OAI inspection ended on Jul 27, 2011. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Five regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
5, 2009–2017
3 OAI · 1 VAI · 1 NAI
Inspections with 483 observations shown
5 of 5
1 posted 483 not matched to an inspection
Form 483s posted by FDA
↗
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    2. Investigation of device failures

      Complaints involving the possible failure of a device and labeling to meet any of its specifications were not investigated where necessary.

    3. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  2. Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Production processes

      Production processes were not controlled to ensure that a device conforms to its specifications.

  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    2. Contact Office Name and Address

      An individual medical device manufacturer report submitted per FDA Form 3500A did not indicate in Block G the device manufacturing site.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Records of complaint investigation

      Records of complaint investigations do not include required information.

    5. Investigation of device failures

      Complaints involving the possible failure of a device and labeling to meet any of its specifications were not investigated where necessary.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by suppliers and contractors have not been adequately established.

    2. Investigation of device failures

      Complaints involving the possible failure of a device and labeling to meet any of its specifications were not evaluated where necessary.

    3. Design validation - user needs and intended uses

      Design validation did not ensure the device conforms to defined user needs and intended uses.

    4. Documentation

      Corrective and preventive action activities and/or results have not been documented.

    5. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    Show 1 more citation
    1. Personnel

      Personnel do not have the necessary training to perform their jobs.

  6. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Procedures

      The procedures addressing documentation of corrective and preventive action activities were not complete.

    2. Supplier oversight

      The type and extent of control to be exercised over services was not clearly defined.

    3. Lack of or inadequate procedures

      The procedures for implementing corrective and preventive actions were not implemented.

    4. Calibration procedures - content

      Calibration procedures do not include specific directions and limits for accuracy and precision.

    5. Investigation

      Certain indicators of nonconformities are not investigated to determine the cause of the nonconformity.

    Show 1 more citation
    1. Design validation - user needs and intended uses

      Design validation did not ensure that devices conform to defined user/patient needs and intended uses.

Regulations FDA cited here

The 9 sections of 21 CFR in FDA’s published citations for this facility’s 5 inspections, with how many of them cited each.

Show all 9
  • 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 5

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Other FEIs recorded as Invacare Corporation

Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1031452 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1031452. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.