Skip to content

Invacare Corporation Form 483, August 2010

FDA issued this Form 483 to Invacare Corporation in Sanford, Florida (FEI 1031452) at the end of a device inspection that ended August 18, 2010. FDA classified the inspection Official Action Indicated (OAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Jan 17, 2024. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Aug 18, 2010
Devices · Compliance: Devices
FDA's outcome
OAI
Official Action Indicated
Posted by FDA
Jan 17, 2024
4,900 days after the inspection
Facility
FEI 1031452
Sanford, Florida

FDA's 6 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. Evaluation of suppliers, contractors, etc., requirements

    Requirements that must be met by suppliers and contractors have not been adequately established.

  2. Investigation of device failures

    Complaints involving the possible failure of a device and labeling to meet any of its specifications were not evaluated where necessary.

  3. Design validation - user needs and intended uses

    Design validation did not ensure the device conforms to defined user needs and intended uses.

  4. Documentation

    Corrective and preventive action activities and/or results have not been documented.

  5. Design Validation - Risk analysis not performed/inadequate

    Risk analysis is incomplete.

  6. Personnel

    Personnel do not have the necessary training to perform their jobs.

After this inspection

FDA's compliance actions for this FEI dated after this inspection and before its next one. FDAgent links a letter when FDA's case number ties it to this facility.

  • Warning LetterDec 15, 2010CDRH · 119 days after this inspection ended

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.