Amsino Healthcare (USA) Inc
330 Corporate Woods Pkwy, Vernon Hills, Illinois, United States
FEI 1417519DUNS 116914622
FDA has 10 inspections on file for Amsino Healthcare (USA) Inc in Vernon Hills, Illinois (FEI 1417519), from 2009 to 2025. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued. Three of the seven FOIA requests for this site have asked FDA for this inspection's 483.
The most recent OAI inspection ended on Dec 19, 2022. No warning letter, court action or import refusal is filed under this FEI.
Seven regulations were cited at more than one inspection; they are listed below.
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, through Dec 31, 2026. Its registrant is Amsino Healthcare (USA) Inc. (D-U-N-S 116914622).
Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 10, 2009–2025
- 2 OAI · 5 VAI · 3 NAI
- Inspections with 483 observations shown
- 4 of 10
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 7 requests, latest FY2023
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
10 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices16 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design plans - Lack of or inadequate
Design plans that describe or reference the design and development activities and define responsibility for implementation have not been established.
Data secured in course of each test
Laboratory records do not include a complete record of all data secured in the course of each test, including all graphs, charts and spectra from laboratory instrumentation, properly identified to show the in-process material, lot tested and drug product tested.
Complaints
Complaint files are not adequately maintained.
Approve or reject components, products
The quality control unit lacks the responsibility and authority to reject all drug products.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
Show 11 more citations
Cleaning / Sanitizing / Maintenance
Equipment and utensils are not cleaned and sanitized at appropriate intervals to prevent contamination that would alter the safety, identity, strength, quality or purity of the drug product.
Design history file
The design history file was not established.
Strict control not exercised over labeling issued
Strict control is not exercised over labeling issued for use in drug product labeling operations.
DMR - not or inadequately maintained
A device master record has not been maintained.
Validation lacking for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile did not include adequate validation of the aseptic process.
Design review - Lack of or inadequate procedures
Procedures for design review have not been established.
Calibration/Inspection/Checking not done
Routine calibration and inspection of mechanical equipment is not performed according to a written program designed to assure proper performance.
Control procedures to monitor and validate performance
Control procedures are not established which monitor the output and validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product.
Investigations of discrepancies, failures
There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Specific information required in individual logs
Individual equipment logs do not show time and lot number of each batch processed.
Cleaning SOP/inspection
Procedures for the cleaning and maintenance of equipment are deficient regarding inspection of the equipment for cleanliness immediately before use.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Procedures for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established and followed.
Show 5 more citations
Management representative
No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.
Cleaning SOPs/instructions
Procedures for the cleaning and maintenance of equipment are deficient regarding sufficient detail of the methods, equipment, and materials used in the cleaning and maintenance operation, and the methods of disassembly and reassembling equipment as necessary to assure proper cleaning and maintenance.
Cleaning / Sanitizing / Maintenance
Equipment and utensils are not cleaned and maintained at appropriate intervals to prevent contamination that would alter the safety, identity, strength, quality or purity of the drug product.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Investigations of discrepancies, failures
There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Calibration/Inspection/Checking not done
Routine inspection and checking of electronic equipment is not performed according to a written program designed to assure proper performance.
Management ensuring quality policy is understood
Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.
Validation lacking for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile do not include adequate validation of the sterilization process.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Show 7 more citations
Cleaning System
Aseptic processing areas are deficient regarding the system for cleaning and disinfecting the equipment to produce aseptic conditions.
Written record of complaint to include findings, follow-up
Written records of investigation of a drug complaint do not include the findings of the investigation and the follow-up.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Air filtration system lacking in production area
The production area air supply lacks an appropriate air filtration system.
Investigations of discrepancies, failures
There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.
Prepared for each batch, include complete information
Batch production and control records do not include complete information relating to the production and control of each batch.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Validation lacking for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile do not include adequate validation of the sterilization process.
Control procedures to monitor and validate performance
Control procedures are not established which validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product.
Implementing Personnel Procedures, Health, Cleanliness.
Requirements addressing the personal practices and clothing of employees have not been implemented.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Seven requests to FDA under the Freedom of Information Act named this site, most recently on FY2023.
Inspection ending Jan 30, 2025VAI
3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Feb 24, 2026. FDAgent holds no copy that FDA posted.
- Dec 31, 2025FDA-FOIA-2025-14426Form 483Records released
- Jul 23, 2025FDA-FOIA-2025-8945Form 483· 1 of 3 sitesClosed
- Jun 18, 2025FDA-FOIA-2025-7776Form 483· 1 of 10 sitesWithdrawn
Inspection ending Nov 13, 2023VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Dec 5, 2023FDA-FOIA-2023-10798Form 483Closed
Inspection ending Dec 19, 2022OAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Oct 2, 2025. FDAgent holds no copy that FDA posted.
- Sep 23, 2025FDA-FOIA-2025-11187Form 483Records released
Inspection date not given
2 requests. None of FDA’s log entries for these requests say records were released.
- FY2023FDA-FOIA-2023-678Form 483Closed
- Oct 13, 2023FDA-FOIA-2023-9008Inspection recordsWithdrawn
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 24 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 10 inspections, with how many of them cited each.
- 21 CFR 211.113Control of microbiological contamination4 of 4
- 21 CFR 211.192Production record review3 of 4
- 21 CFR 820.100Corrective and preventive action3 of 4
- 21 CFR 211.67Equipment cleaning and maintenance2 of 4
- 21 CFR 211.68Automatic, mechanical, and electronic equipment2 of 4
- 21 CFR 211.110Sampling and testing of in-process materials and drug products2 of 4
- 21 CFR 820.20Management responsibility2 of 4
- 21 CFR 211.22Responsibilities of quality control unit1 of 4
Show all 24
- 21 CFR 211.42Design and construction features1 of 4
- 21 CFR 211.46Ventilation, air filtration, air heating and cooling1 of 4
- 21 CFR 211.125Labeling issuance1 of 4
- 21 CFR 211.182Equipment cleaning and use log1 of 4
- 21 CFR 211.188Batch production and control records1 of 4
- 21 CFR 211.194Laboratory records1 of 4
- 21 CFR 211.198Complaint files1 of 4
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 4
- 21 CFR 820.25Personnel1 of 4
- 21 CFR 820.30Design controls1 of 4
- 21 CFR 820.40[Reserved]1 of 4
- 21 CFR 820.70Production and process controls1 of 4
- 21 CFR 820.75Process validation1 of 4
- 21 CFR 820.90Nonconforming product1 of 4
- 21 CFR 820.181Device master record1 of 4
- 21 CFR 820.198Complaint files1 of 4
FDA drug establishment registration
FDA registration search- Registered operations
- Manufacturing
- Registration runs through
- Dec 31, 2026
- Registrant
- Amsino Healthcare (USA) Inc.D-U-N-S 116914622
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1417519 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1417519. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.