Unetixs Vascular, Inc.

333 Strawberry Field Rd Ste 11, Warwick, Rhode Island, United States

FEI 1222117

FDA has 7 inspections on file for Unetixs Vascular, Inc. in Warwick, Rhode Island (FEI 1222117), from 2011 to 2025. The latest ended on and was classified NAI, no action indicated. FDA has issued 3 warning letters under this FEI, the latest on (CDRH).

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two of the four FOIA requests for this site have asked FDA for this inspection's 483.

The most recent OAI inspection ended on Oct 29, 2019. No court action or import refusal is filed under this FEI.

Nine regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
7, 2011–2025
3 OAI · 3 VAI · 1 NAI
Inspections with 483 observations shown
7 of 7
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
4 requests, latest FY2023
Warning letters
· CDRH ↗
3 on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    3. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    4. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    2. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    4. Lack of Written MDR Procedures

      Written MDR procedures have not been developed and maintained.

    5. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    Show 1 more citation
    1. Supplier oversight

      The type and extent of control to be exercised over the product, services, suppliers and contractors was not clearly defined.

  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design control - no procedures

      Procedures for design control have not been established.

    2. Approval, inadequate purchasing data

      Purchasing data that clearly describe or reference specified requirements for purchased or otherwise received product and services have not been adequately established.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    4. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    5. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been established.

    Show 4 more citations
    1. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.

    2. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    3. Procedures not adequately established or maintained

      Document control procedures have not been adequately established.

    4. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities have not been established.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Servicing - Lack of or inadequate procedures

      Procedures or instructions for verifying that servicing meets specified requirements have not been adequately established.

    2. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not reviewed and investigated where necessary.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

  6. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices14 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DMR - not or inadequately maintained

      A device master record has not been maintained.

    2. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been established.

    3. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    4. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    Show 9 more citations
    1. Lack of or inadequate procedures for labeling

      Procedures to control labeling activities have not been established.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    3. Equipment suitability & capability

      Certain inspection, measuring and test equipment is not suitable for its intended purposes and capable of producing valid results.

    4. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to a have not been established.

    5. Documentation of software validation

      Software validation activities and results for computers or automated data processing systems used as part of production have not been documented.

    6. Identification procedures, Lack of or inadequate procedures

      Procedures for identifying product during all stages of receipt, production, distribution, and installation have not been adequately established.

    7. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    8. Design validation - user needs and intended uses

      Design validation did not ensure the device conforms to defined user needs and intended uses.

    9. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

  7. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures for storage.

      Procedures for the control of storage areas and stock rooms have not been adequately established.

    2. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    3. Design control - no procedures

      Procedures for design control have not been established.

    4. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.

    5. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with 21 CFR 820.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Four requests to FDA under the Freedom of Information Act named this site, most recently on FY2023. Each one asked for a Form 483.

  • Inspection ending Dec 4, 2025NAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jun 30, 2026. FDAgent holds no copy that FDA posted.

    1. May 22, 2026FDA-FOIA-2026-5219Form 483Records released
    2. Feb 11, 2026FDA-FOIA-2026-1453Form 483· 1 of 5 sitesOpen
  • Inspection ending Sep 16, 2022VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. FY2023FDA-FOIA-2022-8125Form 483Closed
  • Inspection ending Jul 30, 2021VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. FY2023FDA-FOIA-2023-5306Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 19 sections of 21 CFR in FDA’s published citations for this facility’s 7 inspections, with how many of them cited each.

Show all 19

Warning letters and compliance actions

FDA Data Dashboard compliance actions

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1222117 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1222117. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.