FDA has 5 inspections on file for Verichem Laboratories, Inc. in Providence, Rhode Island (FEI 1221645), from 2010 to 2022. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 6 citations from this inspection: 5 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 806, the device correction and removal reporting rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
The most recent OAI inspection ended on Nov 14, 2014. FDA has issued 2 warning letters under this FEI, the latest on (CDRH). No court action or import refusal is filed under this FEI.
Eight regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 5, 2010–2022
- 2 OAI · 1 VAI · 2 NAI
- Inspections with 483 observations shown
- 4 of 5
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest FY2023
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Justification for not reporting
A justification for not reporting the correction or removal action to FDA that included conclusions, follow-ups and reviews by a designated person was not included in the record.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Show 1 more citation
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Procedures - receipt and dispatch
Procedures were not established for authorizing receipt from and dispatch to storage areas and stock rooms.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Show 4 more citations
Design validation - production units
The design was not validated.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Equipment Installation, Placement, Specified Requirements
The appropriate design, construction, placement, and installation of manufacturing equipment have not been ensured.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.
Statistical techniques - Lack of or inadequate procedures
Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Show 5 more citations
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Lack of or inadequate procedures for handling
Procedures for product handling have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Software validation for automated processes
Software used as part of the quality system has not been adequately validated for its intended use according to an established protocol.
Complaints
Complaint files are not adequately maintained.
Lack of or inadequate procedures for handling
Procedures for product handling have not been adequately established.
Design validation - production units
The design was not validated under defined operating conditions and using initial production units, lots or batches or their equivalents.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Show 3 more citations
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Documentation - review in response to changes or deviations
There is no documentation of the revalidation of a process performed in response to changes or process deviations.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on FY2023. It asked for a Form 483.
Inspection ending May 24, 2022VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- FY2023FDA-FOIA-2023-3261Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 17 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls4 of 4
- 21 CFR 820.70Production and process controls3 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance3 of 4
- 21 CFR 820.198Complaint files3 of 4
- 21 CFR 820.40[Reserved]2 of 4
- 21 CFR 820.50Purchasing controls2 of 4
- 21 CFR 820.75Process validation2 of 4
- 21 CFR 820.140Handling2 of 4
Show all 17
- 21 CFR 820.181Device master record2 of 4
- 21 CFR 806.20Records of corrections and removals not required to be reported1 of 4
- 21 CFR 820.22Quality audit1 of 4
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 4
- 21 CFR 820.90Nonconforming product1 of 4
- 21 CFR 820.100Corrective and preventive action1 of 4
- 21 CFR 820.150Storage1 of 4
- 21 CFR 820.184Device history record1 of 4
- 21 CFR 820.250Statistical techniques1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1221645 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1221645. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.