J&T Health and Wellness LLC
2442 23rd St N, Saint Petersburg, Florida, United States
FEI 1000525872
FDA has 7 inspections on file for J&T Health and Wellness LLC in Saint Petersburg, Florida (FEI 1000525872), from 2011 to 2023. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 8 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.
FDA has issued 2 warning letters under this FEI, the latest on (CDRH). No court action or import refusal is filed under this FEI.
13 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 7, 2011–2023
- 0 OAI · 6 VAI · 1 NAI
- Inspections with 483 observations shown
- 7 of 7
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Jan 7, 2026
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Acceptance activities were not adequately maintained as part of the device history record.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Show 3 more citations
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Documentation of software validation
Software validation activities and results for computers or automated data processing systems used as part of production have not been adequately documented.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality control procedures
You have not established records containing a description of quality control procedures with respect to electronic product radiation safety.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Documentation
Acceptance activities were not adequately documented.
Show 7 more citations
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated and inspected have not been adequately established.
Quality Audits - defined intervals
Quality audits were not performed at defined intervals to determine whether the quality system activities and results comply with quality system procedures.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Lack of System for Determining MDR Events
The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Project Evaluation: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been established.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Show 3 more citations
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed and maintained.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been established.
Lack of Written MDR Procedures
Written MDR procedures have not been maintained.
Show 4 more citations
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been established.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been established.
DMR - not or inadequately maintained
A device master record has not been maintained.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Records not kept
There is no record maintained of a correction or removal action that was not required to be reported to FDA.
Design Validation - Risk analysis not performed/inadequate
Risk analysis was not performed.
- NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety Act2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documented evaluation
The evaluation of potential suppliers was not documented.
Document changes, review and approval, communication
Changes to documents were not reviewed and approved by an individual(s) in the same function or organization that performed the original review and approval.
Design plans - Lack of or inadequate
Design plans that describe or reference the design and development activities and define responsibility for implementation have not been established.
Complaints
Complaint files are not adequately maintained.
Failure to submit
You did not submit an annual report by the September 1 deadline for products requiring one.
Show 5 more citations
Examination for accuracy
Labeling was not sufficiently examined by a designated individual for accuracy including handling instructions and certain additional processing instructions before release.
Training records
Personnel training is not documented.
Management ensuring quality policy is understood
Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.
Timer resetting and causing resumption of radiation emission
The timer was able to automatically reset and cause radiation emission to resume for a period greater than the unused portion of the timer cycle, when emission from the sunlamp product was terminated.
Incoming acceptance records, documentation
Acceptance or rejection of incoming product was not documented.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Jan 7, 2026. Each one asked for a Form 483.
Inspection ending Oct 18, 2023VAI
2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Jan 7, 2026FDA-FOIA-2026-218Form 483Closed
- Feb 12, 2024FDA-FOIA-2024-1507Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 19 sections of 21 CFR in FDA’s published citations for this facility’s 7 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action6 of 7
- 21 CFR 820.80Receiving, in-process, and finished device acceptance5 of 7
- 21 CFR 820.70Production and process controls4 of 7
- 21 CFR 820.72Inspection, measuring, and test equipment4 of 7
- 21 CFR 820.198Complaint files4 of 7
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?3 of 7
- 21 CFR 820.22Quality audit3 of 7
- 21 CFR 820.25Personnel3 of 7
Show all 19
- 21 CFR 820.30Design controls3 of 7
- 21 CFR 820.40[Reserved]3 of 7
- 21 CFR 820.50Purchasing controls3 of 7
- 21 CFR 806.20Records of corrections and removals not required to be reported1 of 7
- 21 CFR 820.20Management responsibility1 of 7
- 21 CFR 820.90Nonconforming product1 of 7
- 21 CFR 820.120Device labeling1 of 7
- 21 CFR 820.181Device master record1 of 7
- 21 CFR 1002.13Annual reports1 of 7
- 21 CFR 1002.30Records to be maintained by manufacturers1 of 7
- 21 CFR 1040.20Sunlamp products and ultraviolet lamps intended for use in sunlamp products1 of 7
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1000525872 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1000525872. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.