Skip to content

21 CFR 1002.30: Records to be maintained by manufacturers

FDA's inspection data lists 38 Form 483 citations of 21 CFR 1002.30 at 29 facilities, from June 2009 to January 24, 2025. 1 warning letter states a violation of it.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
38
At 31 inspections
Facilities cited
29
Distinct FEIs
Most recent citation
Jan 24, 2025
Microlight Corporation of America
Warning letters
1
State a violation of it

The regulation

The current text of 21 CFR 1002.30 from the Electronic Code of Federal Regulations.

§ 1002.30 Records to be maintained by manufacturers.

(a) Manufacturers of products listed under table 1 of § 1002.1 shall establish and maintain the following records with respect to such products:

(1) Description of the quality control procedures with respect to electronic product radiation safety.

(2) Records of the results of tests for electronic product radiation safety, including the control of unnecessary, secondary or leakage electronic product radiation, the methods, devices, and procedures used in such tests, and the basis for selecting such methods, devices, and procedures.

(3) For those products displaying aging effects which may increase electronic product radiation emission, records of the results of tests for durability and stability of the product, and the basis for selecting these tests.

(4) Copies of all written communications between the manufacturer and dealers, distributors, and purchasers concerning radiation safety including complaints, investigations, instructions, or explanations affecting the use, repair, adjustment, maintenance, or testing of the listed product.

(5) Data on production and sales volume levels if available.

(b) In addition to the records required by paragraph (a) of this section, manufacturers of products listed in table 1 of § 1002.1 shall establish and maintain the following records with respect to such products:

(1) A record of the manufacturer's distribution of products in a form which will enable the tracing of specific products or production lots to distributors or to dealers in those instances in which the manufacturer distributes directly to dealers.

(2) Records received from dealers or distributors pursuant to § 1002.41.

21 CFR Part 1002: Records and Reports, Subpart DRead on eCFR

Citations by year

Form 483 citations of 21 CFR 1002.30 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 6 in 2018

Form 483 citations of 21 CFR 1002.30 by inspection year
YearForm 483 citations
20091
20101
20111
20121
20133
20143
20151
20161
20173
20186
20195
20202
20211
20223
20232
20243
20251

How FDA words this citation

FDA's short descriptions and citation text on these 38 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Quality control procedures14 citations
  • Methods, devices, procedures for testing10 citations
  • Results of tests9 citations
  • Basis for selecting methods, devices, procedures3 citations
  • Communication with dealers et. al.1 citations
Show 1 more
  • Received from dealers or distributors1 citations

Citation text

  • “You have not established records containing a description of quality control procedures with respect to electronic product radiation safety.”8 citations
  • “You have not established and maintained records containing a description of quality control procedures with respect to electronic product radiation safety.”4 citations
  • “You have not established and maintained records of the methods, devices and procedures used in tests for electronic product radiation safety.”4 citations
  • “You have not maintained records of the results of tests for electronic product radiation safety, including the control of unnecessary, secondary or leakage electronic product radiation.”4 citations
Show 2 more
  • “You have not established records of the results of tests for electronic product radiation safety, including the control of unnecessary, secondary or leakage electronic product radiation.”3 citations
  • “You have not established records of the methods, devices and procedures used in tests for electronic product radiation safety.”2 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 1002.30, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

Show 4 more

Warning letters stating a violation of 21 CFR 1002.30

1 warning letter FDA posted states a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.

“Failure to conduct certification of an electronic product, based upon a testing program which is in accordance with good manufacturing practices, as required by 21 CFR 1002.30(a)(1).”

Often cited at the same inspection

Of the 31 inspections where FDA cited 21 CFR 1002.30, the share where it also cited each of these sections.

Ask FDAgent about 21 CFR 1002.30

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 1002.30; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.