Microlight Corporation of America
3634 Glenn Lakes Ln Ste 243, Missouri City, Texas, United States
FEI 3003654318
FDA has 2 inspections on file for Microlight Corporation of America in Missouri City, Texas (FEI 3003654318), from 2018 to 2025. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 13 citations from this inspection. They are listed below. One FOIA request has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 2, 2018–2025
- 0 OAI · 2 VAI · 0 NAI
- Inspections with 483 observations shown
- 1 of 2
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Sep 29, 2025
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design review - documentation
The design review results, including identification of the design, the date and the individual(s) performing the review, were not documented in the design history file.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been established.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Design input - documentation
Design input requirements were not documented.
Design history file
The design history file does not demonstrate that the design was developed following the approved design plan.
Show 8 more citations
Acceptable supplier records, inadequate records
Records of acceptable suppliers, contractors and consultants have not been established.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Results of tests
You have not maintained records of the results of tests for electronic product radiation safety, including the control of unnecessary, secondary or leakage electronic product radiation.
Failure to submit
You did not submit an annual report by the September 1 deadline for products requiring one.
Design verification - documentation
The design verification results, including identification of the design, method(s), the date and the individual(s) performing the verification, were not documented in the design history file.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Label to bear a unique device identifier
The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated and investigated where necessary.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Sep 29, 2025. It asked for a Form 483.
Inspection ending Jan 24, 2025VAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Dec 15, 2025. FDAgent holds no copy that FDA posted.
- Sep 29, 2025FDA-FOIA-2025-11403Form 483· 1 of 4 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 9 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 2 inspections.
- 21 CFR 801.20Label to bear a unique device identifier1 of 1
- 21 CFR 820.25Personnel1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.40[Reserved]1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 820.181Device master record1 of 1
- 21 CFR 820.198Complaint files1 of 1
- 21 CFR 1002.13Annual reports1 of 1
Show all 9
- 21 CFR 1002.30Records to be maintained by manufacturers1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3003654318 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3003654318. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.