Empyrean Medical Systems Inc
3010 N Military Trl Ste 220, Boca Raton, Florida, United States
FEI 3023344262
FDA has one inspection on file for Empyrean Medical Systems Inc in Boca Raton, Florida (FEI 3023344262). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 15 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2024
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Oct 21, 2024
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices15 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality control procedures
You have not established records containing a description of quality control procedures with respect to electronic product radiation safety.
Design output - documentation
Design output was not documented before release.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
DMR - not or inadequately maintained
A device master record has not been maintained.
Design verification - Lack of or inadequate procedures
Procedures for design verification have not been adequately established.
Show 10 more citations
Design plans - Lack of or inadequate
Design plans that describe or reference the design and development activities and define responsibility for implementation have not been established.
Complaints
Complaint files are not adequately maintained.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Communication with dealers et. al.
You have not established copies of all written communication between you and dealers, distributors, and purchasers concerning radiation safety, including for your products listed under 21 CFR 1002.1.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Management review - Lack of or inadequate procedures
Procedures for management review have not been adequately established.
Design validation - software validation not performed
Validation of device software was not performed.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Design input - documentation
Design input requirements were not documented.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Oct 21, 2024. Each one asked for a Form 483.
Inspection ending Jun 5, 2024VAI
2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Oct 21, 2024FDA-FOIA-2024-9302Form 483· 1 of 2 sitesClosed
- Jul 17, 2024FDA-FOIA-2024-6297Form 483· 1 of 5 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 9 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 820.20Management responsibility1 of 1
- 21 CFR 820.22Quality audit1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
- 21 CFR 820.181Device master record1 of 1
- 21 CFR 820.184Device history record1 of 1
- 21 CFR 820.198Complaint files1 of 1
Show all 9
- 21 CFR 1002.30Records to be maintained by manufacturers1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3023344262 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3023344262. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.