First Source, Inc.
3495 Winton Pl Bldg E Ste 1, Rochester, New York, United States
FEI 3004063527
FDA has 5 inspections on file for First Source, Inc. in Rochester, New York (FEI 3004063527), from 2016 to 2026. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 3 citations from this inspection. They are listed below. Three of the five FOIA requests for this site have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 820.100(a) and 21 CFR 820.90(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 5, 2016–2026
- 0 OAI · 3 VAI · 2 NAI
- Inspections with 483 observations shown
- 5 of 5
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 5 requests, latest Aug 21, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 8.5.2
Corrective action procedures not documented
Corrective action procedures defining the requirements for evaluating the need for action to ensure that nonconformities do not recur and reviewing the effectiveness of corrective action taken have not been adequately documented.
- ISO 13485:2016 Clause 7.1
Product realization risk management processes not documented
One or more processes for risk management in product realization have not been adequately documented.
- ISO 13485:2016 Clause 7.3.9
Procedures to control design and development changes not documented
Procedures to control design and development changes were not adequately documented.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Tube housing assemblies - cooling curves
Cooling curves for the anode and tube housing were not provided to the purchasers or others upon request, .
Basis for selecting methods, devices, procedures
You have not established and maintained records of the basis for selecting the methods and procedures used in tests for electronic product radiation safety.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Tube housing assemblies (leakage technique factors)
Failure to provide the purchaser or to others upon request at cost, statements of the leakage technique factors for all combinations of tube housing assemblies and beam-limiting devices for which you state compatibility and the peak tube potential at which the aluminum equivalent was obtained.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Show 2 more citations
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Training records
Personnel training is not documented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Supplier notification of changes
There is no agreement with suppliers, contractors and consultants to notify you of changes in the product or service.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Lack of Written MDR Procedures
Written MDR procedures have not been developed and implemented.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Five requests to FDA under the Freedom of Information Act named this site, most recently on Aug 21, 2026. Each one asked for a Form 483.
Inspection ending Jun 24, 2026VAI
3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Aug 26, 2026. FDAgent holds no copy that FDA posted.
- Aug 21, 2026FDA-FOIA-2026-8132Form 483· 1 of 4 sitesOpen
- Jul 31, 2026FDA-FOIA-2026-7331Form 483Records released
- Jul 29, 2026FDA-FOIA-2026-7249Form 483· 1 of 5 sitesRecords released
Inspection ending Sep 19, 2023NAI
2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Oct 30, 2023FDA-FOIA-2023-9530Form 483Closed
- Oct 13, 2023FDA-FOIA-2023-8988Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.50Purchasing controls2 of 5
- 21 CFR 820.90Nonconforming product2 of 5
- 21 CFR 820.100Corrective and preventive action2 of 5
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 5
- 21 CFR 820.20Management responsibility1 of 5
- 21 CFR 820.22Quality audit1 of 5
- 21 CFR 820.25Personnel1 of 5
- 21 CFR 820.30Design controls1 of 5
Show all 12
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 5
- 21 CFR 820.198Complaint files1 of 5
- 21 CFR 1002.30Records to be maintained by manufacturers1 of 5
- 21 CFR 1020.30Diagnostic x-ray systems and their major components1 of 5
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004063527 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004063527. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.