FDA has 2 inspections on file for ITC Nexus Dx in Piscataway, New Jersey (FEI 3005776351), from 2009 to 2010. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 6 citations from this inspection: 4 under 21 CFR Part 820, the device quality system rules, and 2 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.
The most recent OAI inspection ended on Jan 7, 2009. No warning letter, court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 803.50(a)(2) and 21 CFR 820.25(b).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 2, 2009–2010
- 1 OAI · 1 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Acceptance activities were not adequately documented.
Records, DHR {see also 820.184(e)}
Labels and labeling used for each finished product, lot, or batch, were not sufficiently documented in the DHR.
Submission of Supplement to Individual Report Within One Month
A supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.
Design review - Lack of or inadequate procedures
Procedures for design review have not been adequately established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Show 1 more citation
Training records
Personnel training is not documented.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices35 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately and fully validated and approved according to established procedures.
Uniform and timely processing
Complaint handling procedures have not been implemented to ensure that all complaints are processed in a uniform and timely manner.
General
Complaint handling procedures for complaints have not been completed and implemented.
DMR required activities {see also 820.181}
Finished devices were released for distribution prior to completion of activities required in the Device Master Record.
Calibration standard traceability
A calibration standard used for inspection and test equipment is not traceable to national or international standards nor is it an independent reproducible standard.
Show 30 more citations
Changes to correct/prevent quality problems
Changes in methods and procedures needed to correct and prevent identified quality problems are not implemented.
Documentation
Corrective and preventive action activities have not been documented, including the actions needed to correct or prevent recurrence of nonconforming product and other quality problems.
Appropriateness
There is no information to support the appropriateness of the statistical techniques used.
Procedures for product review,disposition lack of/inadequate
Procedures that describe the review and disposition process for nonconforming product were not implemented.
Identification of actions needed
Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.
Design validation - documentation
The results of design validation, including identification of the design, methods, the date, and the individual performing the validation, were not documented and filed in the design history file.
Management review participants
Documentation does not show that management with executive responsibility participated in management reviews.
Specific non-conforming product procedures
Procedures for addressing the identification, documentation, evaluation, segregation, disposition, and investigation of nonconforming product were not complete.
Sampling plans
Sampling plans are not based on valid statistical rationale.
Nature and details of complaint
Records of complaint investigations do not include the nature and details of the complaint.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Acceptance records
The device history record does not include complete acceptance records that demonstrate the device is manufactured in accordance with the device master record.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Quality audits - Lack of or inadequate procedures
Procedures for conducting quality audits were not complete.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication, and evaluation of events that may be subject to medical device reporting requirements.
Design control - no procedures
Procedures to control the design process of the device were not implemented.
Lack of or inadequate procedures
The procedures for implementing corrective and preventive actions were not complete and implemented.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance or rejection of incoming product were not complete and implemented.
Training - Lack of or inadequate procedures
Procedures for identifying training needs were not implemented.
Document review, approval by designated individual
Documents were not reviewed and approved by the individual designated in document control procedures.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not complete and implemented.
Lack of or inadequate final acceptance procedures
Procedures for acceptance or rejection of finished device production runs, lots, or batches were not complete and implemented.
Lack of or inadequate DHR procedures
Procedures have not been implemented to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Design validation - software validation not performed
Validation of device software is incomplete.
Evaluation, timeliness, identification
Complaints representing events that are MDR reportable were not promptly reviewed, evaluated, and investigated by a designated individual.
Incoming inspection, testing, verification
Incoming product was not adequately inspected or tested to verify conformance to specifications.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure that equipment is routinely calibrated were not complete.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Identifying corrective & preventive actions
The procedures addressing identification of corrective and preventive actions were not implemented.
Regulations FDA cited here
The 18 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?2 of 2
- 21 CFR 820.25Personnel2 of 2
- 21 CFR 820.30Design controls2 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 2
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
- 21 CFR 803.56If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?1 of 2
- 21 CFR 820.20Management responsibility1 of 2
- 21 CFR 820.22Quality audit1 of 2
Show all 18
- 21 CFR 820.40[Reserved]1 of 2
- 21 CFR 820.50Purchasing controls1 of 2
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 2
- 21 CFR 820.75Process validation1 of 2
- 21 CFR 820.90Nonconforming product1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.120Device labeling1 of 2
- 21 CFR 820.184Device history record1 of 2
- 21 CFR 820.198Complaint files1 of 2
- 21 CFR 820.250Statistical techniques1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3005776351 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3005776351. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.