20 Corporate Pl S, Piscataway, New Jersey, United States

FEI 3005776351

FDA has 2 inspections on file for ITC Nexus Dx in Piscataway, New Jersey (FEI 3005776351), from 2009 to 2010. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 6 citations from this inspection: 4 under 21 CFR Part 820, the device quality system rules, and 2 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.

The most recent OAI inspection ended on Jan 7, 2009. No warning letter, court action or import refusal is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 803.50(a)(2) and 21 CFR 820.25(b).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
2, 2009–2010
1 OAI · 1 VAI · 0 NAI
Inspections with 483 observations shown
2 of 2
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      Acceptance activities were not adequately documented.

    2. Records, DHR {see also 820.184(e)}

      Labels and labeling used for each finished product, lot, or batch, were not sufficiently documented in the DHR.

    3. Submission of Supplement to Individual Report Within One Month

      A supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.

    4. Design review - Lack of or inadequate procedures

      Procedures for design review have not been adequately established.

    5. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    Show 1 more citation
    1. Training records

      Personnel training is not documented.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices35 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately and fully validated and approved according to established procedures.

    2. Uniform and timely processing

      Complaint handling procedures have not been implemented to ensure that all complaints are processed in a uniform and timely manner.

    3. General

      Complaint handling procedures for complaints have not been completed and implemented.

    4. DMR required activities {see also 820.181}

      Finished devices were released for distribution prior to completion of activities required in the Device Master Record.

    5. Calibration standard traceability

      A calibration standard used for inspection and test equipment is not traceable to national or international standards nor is it an independent reproducible standard.

    Show 30 more citations
    1. Changes to correct/prevent quality problems

      Changes in methods and procedures needed to correct and prevent identified quality problems are not implemented.

    2. Documentation

      Corrective and preventive action activities have not been documented, including the actions needed to correct or prevent recurrence of nonconforming product and other quality problems.

    3. Appropriateness

      There is no information to support the appropriateness of the statistical techniques used.

    4. Procedures for product review,disposition lack of/inadequate

      Procedures that describe the review and disposition process for nonconforming product were not implemented.

    5. Identification of actions needed

      Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.

    6. Design validation - documentation

      The results of design validation, including identification of the design, methods, the date, and the individual performing the validation, were not documented and filed in the design history file.

    7. Management review participants

      Documentation does not show that management with executive responsibility participated in management reviews.

    8. Specific non-conforming product procedures

      Procedures for addressing the identification, documentation, evaluation, segregation, disposition, and investigation of nonconforming product were not complete.

    9. Sampling plans

      Sampling plans are not based on valid statistical rationale.

    10. Nature and details of complaint

      Records of complaint investigations do not include the nature and details of the complaint.

    11. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    12. Acceptance records

      The device history record does not include complete acceptance records that demonstrate the device is manufactured in accordance with the device master record.

    13. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    14. Quality audits - Lack of or inadequate procedures

      Procedures for conducting quality audits were not complete.

    15. Lack of System for Event Evaluations

      The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication, and evaluation of events that may be subject to medical device reporting requirements.

    16. Design control - no procedures

      Procedures to control the design process of the device were not implemented.

    17. Lack of or inadequate procedures

      The procedures for implementing corrective and preventive actions were not complete and implemented.

    18. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance or rejection of incoming product were not complete and implemented.

    19. Training - Lack of or inadequate procedures

      Procedures for identifying training needs were not implemented.

    20. Document review, approval by designated individual

      Documents were not reviewed and approved by the individual designated in document control procedures.

    21. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not complete and implemented.

    22. Lack of or inadequate final acceptance procedures

      Procedures for acceptance or rejection of finished device production runs, lots, or batches were not complete and implemented.

    23. Lack of or inadequate DHR procedures

      Procedures have not been implemented to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.

    24. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    25. Design validation - software validation not performed

      Validation of device software is incomplete.

    26. Evaluation, timeliness, identification

      Complaints representing events that are MDR reportable were not promptly reviewed, evaluated, and investigated by a designated individual.

    27. Incoming inspection, testing, verification

      Incoming product was not adequately inspected or tested to verify conformance to specifications.

    28. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure that equipment is routinely calibrated were not complete.

    29. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    30. Identifying corrective & preventive actions

      The procedures addressing identification of corrective and preventive actions were not implemented.

Regulations FDA cited here

The 18 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.

  • 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?2 of 2
  • 21 CFR 820.25Personnel2 of 2
  • 21 CFR 820.30Design controls2 of 2
  • 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 2
  • 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
  • 21 CFR 803.56If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?1 of 2
  • 21 CFR 820.20Management responsibility1 of 2
  • 21 CFR 820.22Quality audit1 of 2
Show all 18

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3005776351 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3005776351. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.