Med-Mizer, Inc.
80 Commerce Dr, Batesville, Indiana, United States
FEI 3004976058
FDA has 5 inspections on file for Med-Mizer, Inc. in Batesville, Indiana (FEI 3004976058), from 2014 to 2026. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 7 citations from this inspection. They are listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.
The most recent OAI inspection ended on Jul 6, 2015. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
15 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 5, 2014–2026
- 2 OAI · 2 VAI · 1 NAI
- Inspections with 483 observations shown
- 5 of 5
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Sep 16, 2026
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 7.1
Processes needed for product realization not planned and developed
Processes needed for product realization were not adequately planned and developed.
- ISO 13485:2016 Clause 8.2.2
Complaint handling procedures lack required elements
Procedures for timely complaint handling do not adequately include requirements and responsibilities for receiving and recording information and investigating complaints.
- ISO 13485:2016 Clause 6.4.2
Arrangements for control of contaminated product not planned and documented
Arrangements were not adequately planned and documented for the control of contaminated or potentially contaminated product in order to prevent contamination of the work environment, personnel, or product.
- ISO 13485:2016 Clause 8.5.2
Corrective action procedures not documented
Corrective action procedures defining the requirements for evaluating the need for action to ensure that nonconformities do not recur and reviewing the effectiveness of corrective action taken have not been adequately documented.
- ISO 13485:2016 Clause 4.1.2
Risk-based approach not applied to QMS processes
A risk-based approach to the control of the appropriate processes needed for the quality management system has not been adequately applied.
Show 2 more citations
- ISO 13485:2016 Clause 7.5.6
Planned results not consistently met during validation of processes
Validation of processes for production and service provision did not adequately demonstrate the ability of the processes to achieve planned results consistently.
- ISO 13485:2016 Clause 7.5.1
Production and service provision not adequate
Production and service provision was not adequately planned, carried out, monitored, and controlled to ensure that product conforms to specification.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated and maintained have not been adequately established.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Lack of or inadequate DHR procedures
Procedures for device history records have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Show 3 more citations
Sampling methods - Lack of or inadequate procedures
Procedures to ensure sampling methods are adequate for their intended use have not been established.
Lack of or inadequate In-process acceptance procedures
Procedures for the acceptance and control of in-process product have not been adequately established.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Design validation - simulated testing
The design was not validated under actual or simulated use conditions.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices23 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Design input - Lack of or inadequate procedures
Procedures for design input have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated, inspected, checked, and maintained have not been established.
Lack of or inadequate DHR procedures
Procedures for device history records have not been adequately established.
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been established.
Show 18 more citations
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been adequately established.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Examination for accuracy
Labeling was not sufficiently examined by a designated individual for accuracy including handling instructions before release.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been established.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Design output - Lack of or inadequate procedures
Procedures for design output have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Design review - Lack of or inadequate procedures
Procedures for design review have not been adequately established.
Design verification - Lack of or inadequate procedures
Procedures for design verification have not been adequately established.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective identification and evaluation of events that may be subject to medical device reporting requirements.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Management review - Lack of or inadequate procedures
Procedures for management review have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices17 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been established.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Procedures not adequately established or maintained
Document control procedures have not been adequately maintained.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Show 12 more citations
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Design control - no procedures
Procedures for design control have not been established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Lack of Written MDR Procedures
Written MDR procedures have not been maintained and implemented.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate DHR procedures
Procedures for device history records have not been adequately established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Management review - Lack of or inadequate procedures
Procedures for management review have not been established.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Sep 16, 2026. Each one asked for a Form 483.
Inspection ending Jul 31, 2026VAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 30, 2026. FDAgent holds no copy that FDA posted.
- Sep 16, 2026FDA-FOIA-2026-8950Form 483· 1 of 5 sitesRecords released
Inspection ending Aug 6, 2021NAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Mar 19, 2026. FDAgent holds no copy that FDA posted.
- Nov 25, 2025FDA-FOIA-2025-13267Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 18 sections of 21 CFR in FDA’s published citations for this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls4 of 5
- 21 CFR 820.80Receiving, in-process, and finished device acceptance4 of 5
- 21 CFR 820.90Nonconforming product4 of 5
- 21 CFR 820.50Purchasing controls3 of 5
- 21 CFR 820.75Process validation3 of 5
- 21 CFR 820.100Corrective and preventive action3 of 5
- 21 CFR 820.184Device history record3 of 5
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 5
Show all 18
- 21 CFR 820.20Management responsibility2 of 5
- 21 CFR 820.22Quality audit2 of 5
- 21 CFR 820.70Production and process controls2 of 5
- 21 CFR 820.72Inspection, measuring, and test equipment2 of 5
- 21 CFR 820.198Complaint files2 of 5
- 21 CFR 806.10Reports of corrections and removals1 of 5
- 21 CFR 820.25Personnel1 of 5
- 21 CFR 820.40[Reserved]1 of 5
- 21 CFR 820.120Device labeling1 of 5
- 21 CFR 820.250Statistical techniques1 of 5
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004976058 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004976058. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.