Ultroid Technologies, Inc.
3140 W Kennedy Blvd Suite 103, Tampa, Florida, United States
FEI 3004892080
FDA has 4 inspections on file for Ultroid Technologies, Inc. in Tampa, Florida (FEI 3004892080), from 2010 to 2016. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 10 citations from this inspection: 9 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 806, the device correction and removal reporting rules. They are listed below.
The most recent OAI inspection ended on Apr 15, 2015. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
10 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2010–2016
- 1 OAI · 3 VAI · 0 NAI
- Inspections with 483 observations shown
- 4 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Records not kept
There is no record maintained of a correction or removal action that was not required to be reported to FDA.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record and 21 CFR 820.
Management review - Lack of or inadequate procedures
Procedures for management review have not been adequately established.
Show 5 more citations
Availability
Required records are not maintained at a location that is reasonably accessible to responsible officials of the manufacturer and to employees of the FDA and were not made readily available for review and copying by the FDA.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of System for Determining MDR Events
The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.
Availability
Required records are not maintained at a location that is reasonably accessible to responsible officials of the manufacturer and to employees of the FDA and were not made readily available for review and copying by the FDA. Electronic records are used, but they do not meet audit trail and signature manifestation requirements to ensure that they are trustworthy, reliable and generally equivalent to paper records.
Show 3 more citations
Documentation
Corrective and preventive action activities and/or results have not been documented.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Design history file
The design history file was not established and does not demonstrate that the design was developed following the requirements of 21 CFR 820.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices14 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been established.
Show 9 more citations
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Contamination control, Lack of or inadequate procedures
Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Quality System Procedures
Quality system procedures and instructions have not been established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Design history file
The design history file was not established and does not demonstrate that the design was developed following the requirements of 21 CFR 820.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality Audits - defined intervals
Quality audits did not verify that the quality system is effective in fulfilling your quality system objectives.
Quality audit corrective action, reaudits {see also 820.100}
Quality reaudits are not adequately performed.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated and approved according to established procedures.
Determination of relationship of device to event
Investigation of MDR reportable complaints did not include a determination of the relationship of the device to the reported incident or adverse event.
Design transfer - Lack of or inadequate procedures
Procedures were not completed to ensure that the device design was correctly transferred into production.
Show 7 more citations
Examination for accuracy
Labeling was not sufficiently examined by a designated individual for accuracy including the correct expiration date before release.
General
Complaint handling procedures for complaints have not been defined.
Lack of or inadequate procedures
The procedures for implementing corrective and preventive actions were not implemented.
DHR content
The device history record does not demonstrate the device is manufactured in accordance with the device master record.
Lack of or inadequate final acceptance procedures
Procedures for acceptance or rejection of finished device production runs, lots, or batches were not established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Supplier oversight
The type and extent of control to be exercised over the product and suppliers was not clearly defined.
Regulations FDA cited here
The 16 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.22Quality audit4 of 4
- 21 CFR 820.30Design controls4 of 4
- 21 CFR 820.50Purchasing controls4 of 4
- 21 CFR 820.100Corrective and preventive action4 of 4
- 21 CFR 820.184Device history record3 of 4
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 4
- 21 CFR 820.20Management responsibility2 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 4
Show all 16
- 21 CFR 820.180General requirements2 of 4
- 21 CFR 820.198Complaint files2 of 4
- 21 CFR 806.20Records of corrections and removals not required to be reported1 of 4
- 21 CFR 820.40[Reserved]1 of 4
- 21 CFR 820.70Production and process controls1 of 4
- 21 CFR 820.75Process validation1 of 4
- 21 CFR 820.90Nonconforming product1 of 4
- 21 CFR 820.120Device labeling1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004892080 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004892080. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.