Prime Herbs Corporation
1252 Alderwood Ave, Sunnyvale, California, United States
FEI 3004714242
FDA has 4 inspections on file for Prime Herbs Corporation in Sunnyvale, California (FEI 3004714242), from 2012 to 2016. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 8 citations from this inspection: 7 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.
The most recent OAI inspection ended on Nov 6, 2013. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
15 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2012–2016
- 2 OAI · 2 VAI · 0 NAI
- Inspections with 483 observations shown
- 4 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of procedures, or not maintained
Document control procedures have not been established and maintained.
Lack of Written MDR Procedures
Written MDR procedures have not been developed and implemented.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Show 3 more citations
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been established.
Quality System Procedures
Quality system procedures and instructions have not been established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Supplier oversight
The type and extent of control to be exercised over the product and contractors was not clearly defined.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with 21 CFR 820.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by contractors have not been adequately established.
Show 5 more citations
Personnel
Personnel do not have the necessary education, training, and experience to perform their jobs.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design control - no procedures
Procedures for design control have not been established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
DHR documentation of label release {see also 820.184}
The DHR does not include records of examination and release of device labeling, including date and signature of the examiner.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Lack of or inadequate procedures for labeling
Procedures to control labeling activities have not been adequately established.
Design output - review and approval
Design output was not reviewed before release.
Show 5 more citations
Personnel
Personnel do not have the necessary education, background, training, and experience to perform their jobs.
Supplier oversight
The type and extent of control to be exercised over the product and services was not clearly defined.
Availability
Required records are not maintained at a location that is reasonably accessible to responsible officials of the manufacturer and to employees of the FDA.
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by contractors have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices16 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design control - no procedures
Procedures for design control have not been established.
Lack of or inadequate DHR procedures
Procedures for device history records have not been established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
DMR - not or inadequately maintained
A device master record has not been maintained.
Lack of procedures, or not maintained
Document control procedures have not been established.
Show 11 more citations
Personnel
Personnel do not have the necessary training and experience to perform their jobs.
Packaging
Device packaging and/or shipping containers are not designed and constructed to protect the device from alteration or damage during processing, storage, handling, and distribution.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been established.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed, maintained, and implemented.
Quality System Procedures
Quality system procedures and instructions have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Lack of or inadequate procedures for distribution
Procedures for control and distribution of finished devices have not been established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
Lack of or inadequate procedures for labeling
Procedures to control labeling activities have not been established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Regulations FDA cited here
The 18 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.25Personnel4 of 4
- 21 CFR 820.50Purchasing controls4 of 4
- 21 CFR 820.20Management responsibility3 of 4
- 21 CFR 820.30Design controls3 of 4
- 21 CFR 820.100Corrective and preventive action3 of 4
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 4
- 21 CFR 820.22Quality audit2 of 4
- 21 CFR 820.40[Reserved]2 of 4
Show all 18
- 21 CFR 820.75Process validation2 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 4
- 21 CFR 820.120Device labeling2 of 4
- 21 CFR 820.184Device history record2 of 4
- 21 CFR 820.198Complaint files2 of 4
- 21 CFR 820.90Nonconforming product1 of 4
- 21 CFR 820.130Device packaging1 of 4
- 21 CFR 820.160Distribution1 of 4
- 21 CFR 820.180General requirements1 of 4
- 21 CFR 820.181Device master record1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004714242 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004714242. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.