LifeSouth Community Blood Centers, Inc
4039 Newberry Rd, Gainesville, Florida, United States
FEI 3003707120DUNS 085091726
FDA has 12 inspections on file for LifeSouth Community Blood Centers, Inc in Gainesville, Florida (FEI 3003707120), from 2010 to 2026. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 2 citations from this inspection, all under 21 CFR Part 211, the manufacturing rules for finished drugs. They are listed below. Two of the four FOIA requests for this site have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Five regulations were cited at more than one inspection; they are listed below.
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, packing, labeling and analysis, through Dec 31, 2027. Its registrant is LifeSouth Community Blood Centers, Inc. (D-U-N-S 085091726).
Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 12, 2010–2026
- 0 OAI · 6 VAI · 6 NAI
- Inspections with 483 observations shown
- 5 of 12
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 4 requests, latest Oct 21, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
12 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Cleaning SOPs/instructions
Procedures for the cleaning and maintenance of equipment are deficient regarding sufficient detail of the methods, equipment, and materials used in the cleaning and maintenance operation, and the methods of disassembly and reassembling equipment as necessary to assure proper cleaning and maintenance.
Cleaning / Sanitizing / Maintenance
Equipment and utensils are not cleaned and maintained at appropriate intervals to prevent contamination that would alter the safety, identity, strength, quality or purity of the drug product.
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Management review - Lack of or inadequate procedures
Procedures for management review have not been established.
Design validation - Risk analysis
Results of the design risk analysis were not documented.
Management representative
No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.
Documentation of software validation
Software validation activities and results for computers or automated data processing systems used as part of the quality system have not been adequately documented.
Show 5 more citations
Documentation
Corrective and preventive action activities and/or results have not been documented.
Lack of System for Determining MDR Events
The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design validation - software validation documentation
Results of the validation of the device software were not adequately documented.
Personnel
Personnel do not have the necessary training to perform their jobs.
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene TherapiesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Authority lacking to review records, investigate errors
The quality control unit lacks authority to review production records to assure that no errors have occurred.
Environmental Monitoring System
Aseptic processing areas are deficient regarding the system for monitoring environmental conditions.
When to report
Biological product deviations were not reported within the 45 calendar day timeframe.
Procedures not in writing, fully followed
The responsibilities and procedures applicable to the quality control unit are not in writing.
Computer control of master formula records
Appropriate controls are not exercised over computers or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel.
Show 7 more citations
Cleaning System
Aseptic processing areas are deficient regarding the system for cleaning and disinfecting the room and equipment to produce aseptic conditions.
Package label - small containers
The container label is not capable of bearing a full label, but the package label does not include the address and license number of the manufacturer and the expiration date.
Investigations of discrepancies, failures
There is a failure to thoroughly review any unexplained discrepancy whether or not the batch has been already distributed.
Scientifically sound laboratory controls
Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity.
Master production and control records, procedure
The preparation of your master production and control records was not described in a written procedure.
Reserve drug product sample quantity - all tests
The reserve sample of drug product does not consist of at least twice the quantity necessary to perform all the required tests of drug product.
Review of representative number of batches
Written procedures are not followed for evaluations conducted at least annually to review records associated with a representative number of batches, whether approved or rejected.
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene TherapiesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design control - no procedures
Procedures for design control have not been established.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene Therapies16 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been established.
Thorough investigations
Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy and a failure of a lot or unit to meet any of its specifications.
Person performing, test results, interpretation
Records fail to show test results so as to provide a complete history of the work performed.
Major changes to an approved application
Failure to submit a supplement prior to distributing product made using the change.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Show 11 more citations
Storage temps./agitation
Failure to store platelets immediately after resuspension at 20 to 24 degrees Celsius with continuous gentle agitation.
Biological product deviation report
Failure to submit a biological product deviation report.
Acceptance status
The acceptance status of product was not maintained.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Design review - documentation
The design review results, including , were not documented in the design history file.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Validation of changes to automated process software
Changes to software used as part of production and the quality system were not adequately validated before approval and issuance.
Lack of Written MDR Procedures
Written MDR procedures have not been developed, maintained, and implemented.
Software validation for automated processes
Software used as part of production and the quality system has not been adequately validated for its intended use according to an established protocol.
Required records
Failure to maintain donor records.
Quality System Procedures
Quality system procedures and instructions have not been established.
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Four requests to FDA under the Freedom of Information Act named this site, most recently on Oct 21, 2026. Each one asked for a Form 483.
Inspection ending May 21, 2026VAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 17, 2026. FDAgent holds no copy that FDA posted.
- Oct 21, 2026FDA-FOIA-2026-10213Form 483· 1 of 4 sitesRecords released
- Jul 14, 2026FDA-FOIA-2026-6674Form 483· 1 of 4 sitesRecords released
Inspection ending Nov 9, 2023VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Jan 30, 2024FDA-FOIA-2024-1008Form 483Closed
Inspection date not given
1 request. None of FDA’s log entries for these requests say records were released.
- May 20, 2024FDA-FOIA-2024-4497Form 483 + EIRClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 25 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 12 inspections, with how many of them cited each.
- 21 CFR 820.22Quality audit3 of 5
- 21 CFR 820.30Design controls3 of 5
- 21 CFR 820.100Corrective and preventive action3 of 5
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 5
- 21 CFR 820.20Management responsibility2 of 5
- 21 CFR 820.70Production and process controls2 of 5
- 21 CFR 211.22Responsibilities of quality control unit1 of 5
- 21 CFR 211.42Design and construction features1 of 5
Show all 25
- 21 CFR 211.67Equipment cleaning and maintenance1 of 5
- 21 CFR 211.68Automatic, mechanical, and electronic equipment1 of 5
- 21 CFR 211.160General requirements1 of 5
- 21 CFR 211.170Reserve samples1 of 5
- 21 CFR 211.180General requirements1 of 5
- 21 CFR 211.186Master production and control records1 of 5
- 21 CFR 211.192Production record review1 of 5
- 21 CFR 600.14Reporting of biological product deviations by licensed manufacturers1 of 5
- 21 CFR 606.100Standard operating procedures1 of 5
- 21 CFR 606.160Records1 of 5
- 21 CFR 606.171Reporting of product deviations by licensed manufacturers, unlicensed registered blood establishments, and transfusion services1 of 5
- 21 CFR 640.25General requirements1 of 5
- 21 CFR 820.25Personnel1 of 5
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 5
- 21 CFR 820.86Acceptance status1 of 5
- 21 CFR 601.12Changes to an approved application1 of 5
- 21 CFR 610.60Container label1 of 5
FDA drug establishment registration
FDA registration search- Registered operations
- Manufacturing, packing, labeling, analysis
- Registration runs through
- Dec 31, 2027
- Registrant
- LifeSouth Community Blood Centers, Inc.D-U-N-S 085091726
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Other FEIs recorded as LifeSouth Community Blood Centers, Inc
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
LifeSouth Community Blood Centers, IncFEI 3001451318Gainesville, Georgia, United States
Last inspected 2026-08-12NAI
LifeSouth Community Blood Centers, IncFEI 3019562580Newberry, Florida, United States
Last inspected 2022-09-08NAI
LifeSouth Community Blood Centers, IncFEI 3013694007Cottondale, Alabama, United States
Last inspected 2021-08-23NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3003707120 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3003707120. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.