Skip to content

JAS Diagnostics, Inc./Drew Scientific, Inc.

14050 Nw 57th Ct, Miami Lakes, Florida, United States

FEI 3003071743

FDA has 6 inspections on file for JAS Diagnostics, Inc./Drew Scientific, Inc. in Miami Lakes, Florida (FEI 3003071743), from 2009 to 2025. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 5 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.

FDA issued a warning letter on (CDRH). 1 import refusal is filed under this FEI, on . No court action is filed under this FEI.

Eight regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
6, 2009–2025
2 OAI · 3 VAI · 1 NAI
Inspections with 483 observations shown
5 of 6
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Mar 23, 2026
Import refusals
1 refusal
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design control - no procedures

      Procedures for design control have not been established.

    2. Training records

      Personnel training is not documented.

    3. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    4. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

    5. Complaints

      Complaint files are not maintained.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities have not been adequately established.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    5. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    Show 4 more citations
    1. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    3. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    4. Design control - no procedures

      Procedures for design control have not been established.

  3. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    3. Documentation

      Process validation activities and results have not been adequately documented.

    4. Lack of or inadequate organizational structure

      The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.

    5. Design history file

      The design history file was not established.

    Show 3 more citations
    1. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Design review - Lack of or inadequate procedures

      Procedures for design review have not been adequately established.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    3. Packaging

      Device packaging and/or shipping containers are not designed and constructed to protect the device from alteration or damage during processing, storage, handling, and distribution.

    4. Training records

      Personnel training is not documented.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    Show 5 more citations
    1. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

    2. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    3. Design history file

      The design history file was not established.

    4. Lack of System for Timely Submission of Reports

      The written MDR procedure does not include an internal system which provides for timely transmission of complete medical device reports to FDA.

    5. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by suppliers, contractors, and consultants have not been adequately established.

  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
  6. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Servicing - Lack of or inadequate procedures

      Servicing procedures have not been established.

    2. Packaging

      Device packaging and/or shipping containers are not designed and constructed well enough to protect the device from damage during processing, storage, handling, and distribution.

    3. Documentation of validated process performance

      There is no documentation of the date performed, the individual performing the process, and the major equipment used for a validated process.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Mar 23, 2026. Each one asked for a Form 483.

  • Inspection ending Jun 12, 2025OAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 11, 2025. FDAgent holds no copy that FDA posted.

    1. Mar 23, 2026FDA-FOIA-2026-2864Form 483Records released
    2. Aug 5, 2025FDA-FOIA-2025-9488Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 15 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 6 inspections, with how many of them cited each.

Show all 15

Warning letters and compliance actions

FDA Data Dashboard compliance actions

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3003071743, one row per refused product line.

1 refusal on 2017-06-07 · charged FRNMFGREG on 1, REGISTERED on 1 · no FDA sample analysis

FDA firm profile
  • 2017-06-07

    GENERAL PURPOSE REAGENT88PPM

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-0732706-5/11/1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3003071743 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3003071743. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.