Medefil Incorporated

250 Windy Point Dr, Glendale Heights, Illinois, United States

FEI 3001677091DUNS 080738829

FDA has 11 inspections on file for Medefil Incorporated in Glendale Heights, Illinois (FEI 3001677091), from 2009 to 2025. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued. Five FOIA requests have asked FDA for this inspection's 483.

The most recent OAI inspection ended on Feb 11, 2010. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Six regulations were cited at more than one inspection; they are listed below.

As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, packing, labeling, analysis, sterilization, relabeling and repacking, through Dec 31, 2027. Its registrant is Medefil, LLC (D-U-N-S 016448669).

Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
11, 2009–2025
1 OAI · 7 VAI · 3 NAI
Inspections with 483 observations shown
6 of 11
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
5 requests, latest May 15, 2026
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Drug establishment registration
Through

Inspection history

11 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
  2. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been established to control product that does not conform to specified requirements.

    2. Procedures To Be in Writing

      Written procedures are lacking which describe in sufficient detail the sampling, testing, approval and rejection of components and drug product containers.

    3. Containers & Closures Clean, Sterilized, Pyrogen-free

      Drug product containers were not clean and sterilized and processed to remove pyrogenic properties to assure that they are suitable for their intended use.

    4. Contamination control, Lack of or inadequate procedures

      Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been adequately established.

    5. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    Show 2 more citations
    1. Procedures for sterile drug products

      Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established.

    2. Procedures not in writing, fully followed

      The responsibilities and procedures applicable to the quality control unit are not fully followed.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Procedures for sterile drug products

      Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established, written, and followed.

    2. Quality audits - auditor independence

      Individuals who conduct quality audits have direct responsibility for the matters being audited.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Investigations of discrepancies, failures

      There is a failure to thoroughly review any unexplained discrepancy whether or not the batch has been already distributed.

    5. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documented evaluation

      The evaluation of potential suppliers, contractors, and consultants was not documented.

    2. Lack of System for Event Evaluations

      The written MDR Procedure does not include an internal system which provides for the timely and effective identification and evaluation of events that may be subject to medical device reporting requirements.

  6. NAINo objectionable conditions that need follow-upDrugs · Bioresearch MonitoringNone
  7. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of System for Event Evaluations

      The written MDR Procedure does not include an internal system which provides for the timely and effective identification of events that may be subject to medical device reporting requirements.

    2. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    3. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    4. Scientifically sound laboratory controls

      Laboratory controls do not include the establishment of scientifically sound and appropriate specifications and test procedures designed to assure that components, drug product containers, closures, and drug products conform to appropriate standards of identity, strength, quality and purity.

    5. Deviations from laboratory control requirements

      Deviations from written standards, test procedures, and laboratory mechanisms are not recorded and justified.

  8. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  9. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      Process validation activities have not been adequately documented.

    2. Design validation - user needs and intended uses

      Design validation did not ensure the device conforms to defined user needs and intended uses.

    3. Design input - documentation

      Design input requirements were not adequately documented.

    4. Documentation

      Corrective and preventive action activities and/or results have not been documented.

    5. Design validation - Risk analysis

      Results of the design risk analysis were not adequately documented.

    Show 3 more citations
    1. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate and is incomplete.

    2. Design output - documentation

      Design output was not adequately documented before release.

    3. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

  10. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
  11. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    2. Evaluation of suppliers, contractors, etc., requirements

      Adequate quality requirements that must be met by suppliers were not established and implemented.

    3. Documented evaluation

      The evaluation of potential suppliers was not documented.

    4. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not established and implemented.

    5. Investigation procedures

      The corrective and preventive action procedures addressing the investigation of the cause of nonconformities relating to product, processes, and the quality system were not implemented.

    Show 1 more citation
    1. Complete test data included in records

      Laboratory records do not include complete data derived from all tests, examinations and assay necessary to assure compliance with established specifications and standards.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Five requests to FDA under the Freedom of Information Act named this site, most recently on May 15, 2026. Each one asked for a Form 483.

  • Inspection ending Apr 21, 2025VAI

    5 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jul 15, 2025. FDAgent holds no copy that FDA posted.

    1. May 15, 2026FDA-FOIA-2026-4993Form 483· 1 of 5 sitesWithdrawn
    2. May 14, 2026FDA-FOIA-2026-4934Form 483· 1 of 5 sitesWithdrawn
    3. Jul 31, 2025FDA-FOIA-2025-9258Form 483Closed
    4. Jun 30, 2025FDA-FOIA-2025-8109Form 483· 1 of 5 sitesRecords released
    5. Jun 23, 2025FDA-FOIA-2025-7938Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 17 sections of 21 CFR in FDA’s published citations for 6 of this facility’s 11 inspections, with how many of them cited each.

Show all 17

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

FDA drug establishment registration

FDA registration search
Registered operations
Manufacturing, packing, labeling, analysis, sterilization, relabeling, repacking
Registration runs through
Dec 31, 2027
Registrant
Medefil, LLCD-U-N-S 016448669

From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3001677091 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3001677091. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.