TriMed Inc.

27533 Avenue Hopkins, Valencia, California, United States

FEI 3001236812

FDA has 6 inspections on file for TriMed Inc. in Valencia, California (FEI 3001236812), from 2012 to 2024. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

The most recent OAI inspection ended on Mar 18, 2016. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

10 regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
6, 2012–2024
1 OAI · 3 VAI · 2 NAI
Inspections with 483 observations shown
5 of 6
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design verification - documentation

      The design verification results, including identification of the design, method(s), the date and the individual(s) performing the verification, were not adequately documented in the design history file.

    2. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

  2. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  3. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design control - no procedures

      Procedures for design control have not been established.

    2. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    3. Documentation

      Acceptance activities were not adequately documented.

    4. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    Show 1 more citation
    1. Lack of or inadequate DHR procedures

      Procedures for device history records have not been adequately established.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    2. Lack of or inadequate procedures for labeling

      Procedures to control labeling activities have not been adequately established.

    3. Investigation of device failures

      Complaints involving the possible failure of a device, labeling and packaging to meet any of its specifications were not investigated where necessary.

    4. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

    5. Software validation for automated processes

      Software used as part of the quality system has not been validated for its intended use according to an established protocol.

    Show 6 more citations
    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    2. Lack of or inadequate DHR procedures

      Procedures for device history records have not been adequately established.

    3. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    5. Lack of procedures, or not maintained

      Document control procedures have not been established and maintained.

    6. Design control - no procedures

      Procedures for design control have not been established.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices13 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    2. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    3. Records not kept

      There is no record maintained of a correction or removal action that was not required to be reported to FDA.

    4. Lack of or inadequate DHR procedures

      Procedures for device history records have not been adequately established.

    5. Investigation of device failures

      Complaints involving the possible failure of a device , labeling and packaging to meet any of its specifications were not reviewed , evaluated and investigated where necessary.

    Show 8 more citations
    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Lack of or inadequate procedures for handling

      Procedures for product handling have not been adequately established.

    3. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    4. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    5. Lack of Written MDR Procedures

      Written MDR procedures have not been implemented.

    6. Software validation for automated processes

      Software used as part of the quality system has not been validated for its intended use according to an established protocol.

    7. Design control - no procedures

      Procedures for design control have not been established.

    8. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been established.

  6. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Equipment control activity documentation

      Equipment calibrations have not been documented.

    2. Investigation of device failures

      Complaints involving the possible failure of a device, labeling, and packaging to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.

    3. Design history file

      The design history file was not established.

    4. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    5. Quality Audits - defined intervals

      Quality audits were not performed at defined intervals to determine whether the quality system activities and results comply with quality system procedures.

Regulations FDA cited here

The 15 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 6 inspections, with how many of them cited each.

Show all 15

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3001236812 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3001236812. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.