TriMed Inc.
27533 Avenue Hopkins, Valencia, California, United States
FEI 3001236812
FDA has 6 inspections on file for TriMed Inc. in Valencia, California (FEI 3001236812), from 2012 to 2024. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
The most recent OAI inspection ended on Mar 18, 2016. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
10 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 6, 2012–2024
- 1 OAI · 3 VAI · 2 NAI
- Inspections with 483 observations shown
- 5 of 6
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design verification - documentation
The design verification results, including identification of the design, method(s), the date and the individual(s) performing the verification, were not adequately documented in the design history file.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design control - no procedures
Procedures for design control have not been established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Documentation
Acceptance activities were not adequately documented.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Show 1 more citation
Lack of or inadequate DHR procedures
Procedures for device history records have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Lack of or inadequate procedures for labeling
Procedures to control labeling activities have not been adequately established.
Investigation of device failures
Complaints involving the possible failure of a device, labeling and packaging to meet any of its specifications were not investigated where necessary.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Software validation for automated processes
Software used as part of the quality system has not been validated for its intended use according to an established protocol.
Show 6 more citations
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of or inadequate DHR procedures
Procedures for device history records have not been adequately established.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of procedures, or not maintained
Document control procedures have not been established and maintained.
Design control - no procedures
Procedures for design control have not been established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Records not kept
There is no record maintained of a correction or removal action that was not required to be reported to FDA.
Lack of or inadequate DHR procedures
Procedures for device history records have not been adequately established.
Investigation of device failures
Complaints involving the possible failure of a device , labeling and packaging to meet any of its specifications were not reviewed , evaluated and investigated where necessary.
Show 8 more citations
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate procedures for handling
Procedures for product handling have not been adequately established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Lack of Written MDR Procedures
Written MDR procedures have not been implemented.
Software validation for automated processes
Software used as part of the quality system has not been validated for its intended use according to an established protocol.
Design control - no procedures
Procedures for design control have not been established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Equipment control activity documentation
Equipment calibrations have not been documented.
Investigation of device failures
Complaints involving the possible failure of a device, labeling, and packaging to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.
Design history file
The design history file was not established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Quality Audits - defined intervals
Quality audits were not performed at defined intervals to determine whether the quality system activities and results comply with quality system procedures.
Regulations FDA cited here
The 15 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls5 of 5
- 21 CFR 820.198Complaint files4 of 5
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?3 of 5
- 21 CFR 820.80Receiving, in-process, and finished device acceptance3 of 5
- 21 CFR 820.100Corrective and preventive action3 of 5
- 21 CFR 820.184Device history record3 of 5
- 21 CFR 820.22Quality audit2 of 5
- 21 CFR 820.70Production and process controls2 of 5
Show all 15
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 5
- 21 CFR 806.20Records of corrections and removals not required to be reported1 of 5
- 21 CFR 820.40[Reserved]1 of 5
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 5
- 21 CFR 820.90Nonconforming product1 of 5
- 21 CFR 820.120Device labeling1 of 5
- 21 CFR 820.140Handling1 of 5
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3001236812 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3001236812. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.