Zynex Medical, Inc.

8181 E Tufts Ave Ste 200, Denver, Colorado, United States

FEI 3000214920

FDA has 6 inspections on file for Zynex Medical, Inc. in Denver, Colorado (FEI 3000214920), from 2009 to 2022. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

The most recent OAI inspection ended on Aug 8, 2017. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

10 regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
6, 2009–2022
2 OAI · 2 VAI · 2 NAI
Inspections with 483 observations shown
4 of 6
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    2. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    3. Design validation - software validation not performed

      Validation of device software was not performed.

    4. Product rework documentation, DHR {see also 820.184}

      Rework and reevaluation activities have not been fully documented in the device history record.

    5. Design validation - Risk analysis

      Results of the design risk analysis were not adequately documented.

    Show 7 more citations
    1. Design validation - documentation

      The results of design validation, including identification of the design, method(s), the date and the individual(s) performing validation, were not documented in the design history file.

    2. Training records

      Personnel training is not documented.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    4. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    5. Analyzing service report

      Service reports were not analyzed following appropriate statistical methods.

    6. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities have not been adequately established.

    7. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Product rework documentation, DHR {see also 820.184}

      Rework and reevaluation activities have not been fully documented in the device history record.

    3. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    4. Design validation - Risk analysis

      Results of the design risk analysis were not adequately documented.

    5. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    Show 2 more citations
    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    2. Design validation - documentation

      The results of design validation, including identification of the design, method(s), the date, and the individual(s) performing validation, were not documented in the design history file.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices19 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been established.

    2. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been established.

    3. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities have not been adequately established.

    4. Complaints

      Complaint files are not maintained.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    Show 14 more citations
    1. Design input - Lack of or inadequate procedures

      Procedures for design input have not been adequately established.

    2. Evaluation and Selection, Suppliers, Contractors, etc.

      Potential suppliers were not evaluated and selected based on their ability to meet specified requirements.

    3. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    4. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    5. Design verification - Lack of or inadequate procedures

      Procedures for design verification have not been adequately established.

    6. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    7. Design validation - Risk analysis

      Results of the design risk analysis were not adequately documented.

    8. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been established.

    9. Production processes

      Production processes were not to ensure that a device conforms to its specifications.

    10. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    11. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    12. Lack of or inadequate DHR procedures

      Procedures for device history records have not been established.

    13. Product rework procedures, Lack of or inadequate procedures

      Procedures for rework of nonconforming product have not been established.

    14. Design validation - production units

      The design was not validated.

  6. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate DHR procedures

      Procedures have not been established to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.

    2. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    3. Design Validation - Risk analysis not performed/inadequate

      Risk analysis was not performed.

    4. Quality audits - Lack of or inadequate procedures

      Procedures for conducting quality audits were not complete.

    5. Design review - Lack of or inadequate procedures

      Procedures for planning and conducting reviews of the design results at appropriate stages of the device's design development were not established.

    Show 1 more citation
    1. Analysis of data/reports from data sources

      Not all data from quality data sources are analyzed to identify existing and potential causes of nonconforming product and other quality problems.

Regulations FDA cited here

The 14 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 6 inspections, with how many of them cited each.

Show all 14

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3000214920 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3000214920. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.