Abbott Medical

6901 Preston Rd, Plano, Texas, United States

FEI 1627487

FDA has 11 inspections on file for Abbott Medical in Plano, Texas (FEI 1627487), from 2009 to 2025. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

The most recent OAI inspection ended on May 2, 2011. FDA has issued 2 warning letters under this FEI, the latest on (CDRH). No court action or import refusal is filed under this FEI.

Six regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
11, 2009–2025
3 OAI · 1 VAI · 7 NAI
Inspections with 483 observations shown
4 of 11
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Sep 28, 2021
Warning letters
· CDRH ↗
2 on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

11 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Project Evaluation: DevicesNone
  2. NAINo objectionable conditions that need follow-upDevices · Project Evaluation: DevicesNone
  3. NAINo objectionable conditions that need follow-upDevices · Project Evaluation: DevicesNone
  4. NAINo objectionable conditions that need follow-upDevices · Project Evaluation: DevicesNone
  5. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  6. NAINo objectionable conditions that need follow-upDevices · Bioresearch MonitoringNone
  7. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
  8. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Procedures not adequately established or maintained

      Document control procedures have not been adequately established and maintained.

  9. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Design verification - output does not meet input requirement

      Design verification does not confirm that design output meets design input requirements.

    3. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    4. Incorrect translation to production specifications

      The device design was not correctly translated into production specifications.

    5. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    Show 4 more citations
    1. Design validation - user needs and intended uses

      Design validation did not ensure the device conforms to defined user needs and intended uses.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures for monitoring and control of process parameters for a validated process have not been established.

    4. Documentation

      Acceptance activities were not documented and maintained as part of the device history record.

  10. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    3. Report of violation of the Act (see 803.52(e)(9))

      A violation of the FD&C Act involving a device which might present a risk to health was not reported to FDA.

    4. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    5. Submission of Supplement to Individual Report Within One Month

      A supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.

    Show 6 more citations
    1. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    2. Process changes - review, evaluation and revalidation

      A validated process was not revalidated when changes or process deviations occurred.

    3. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been established to control product that does not conform to specified requirements.

    4. Design verification - output does not meet input requirement

      Design verification does not confirm that design output meets design input requirements.

    5. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    6. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  11. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      The procedures for implementing corrective and preventive actions were not complete and implemented.

    2. Identifying corrective & preventive actions

      The procedures addressing identification of corrective and preventive actions were not implemented.

    3. Lack of or inadequate final acceptance procedures

      Procedures for acceptance or rejection of finished device production runs, lots, or batches were not complete.

    4. Quality Audit/Reaudit - conducted

      Quality audits were not conducted at sufficient regular intervals, as prescribed by internal procedures to verify that the quality system is effective in fulfilling your quality system objectives.

    5. Identification of actions needed

      Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.

    Show 4 more citations
    1. General

      Complaint handling procedures for receiving and evaluating complaints have not been completed.

    2. Investigation

      Certain indicators of nonconformities are not investigated to determine the cause of the nonconformity.

    3. CAPA verification/validation of corrective/preventive action

      Corrective and preventive actions have not been verified or validated to ensure that the action is effective and does not adversely affect the finished device.

    4. Design validation - user needs and intended uses

      Design validation did not ensure that devices conform to defined user/patient needs and intended uses.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Sep 28, 2021. It asked for a Form 483.

  • Inspection ending Sep 15, 2017

    1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.

    1. Sep 28, 2021FDA-FOIA-2021-6450Form 483No records found

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 12 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 11 inspections, with how many of them cited each.

Show all 12

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.

Other FEIs recorded as Abbott Medical

Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.

  • Abbott MedicalFEI 3007113487Saint Paul, Minnesota, United States

    Last inspected 2025-02-21NAI

  • Abbott MedicalFEI 2017865Sylmar, California, United States

    Last inspected 2023-06-28VAI

  • Abbott MedicalFEI 3006705815Arecibo, Puerto Rico, United States

    Last inspected 2022-11-23VAI

  • Abbott MedicalFEI 3008377825Penang, Malaysia

    Last inspected 2018-05-11NAI

  • Abbott MedicalFEI 2648612Caguas, Puerto Rico, United States

    Last inspected 2016-02-17NAI

  • Abbott MedicalFEI 2938836Sunnyvale, California, United States

    Last inspected 2016-01-22NAI

  • Abbott MedicalFEI 3009600098Westford, Massachusetts, United States

    Last inspected 2014-01-29NAI

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1627487 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1627487. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.