Abbott Medical
Lot A Interior -Carr #2 Km 67.5, Santana Indl. Park, Sector Los Pinos, Arecibo, Puerto Rico, United States
FEI 3006705815
FDA has 8 inspections on file for Abbott Medical in Arecibo, Puerto Rico (FEI 3006705815), from 2010 to 2022. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published one citation from this inspection, under 21 CFR Part 820, the device quality system rules. It is listed below.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 8, 2010–2022
- 0 OAI · 5 VAI · 3 NAI
- Inspections with 483 observations shown
- 3 of 8
- 1 posted 483 not matched to an inspection
- Form 483s posted by FDA
- ↗
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
- NAINo objectionable conditions that need follow-upDevices · Project Evaluation: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- NAINo objectionable conditions that need follow-upDevices · Project Evaluation: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Project Evaluation: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Supplier notification of changes
There is no agreement with contractors to notify you of changes in the product or service.
Documentation
Process validation results have not been adequately documented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Project Evaluation: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Supplier oversight
The type and extent of control to be exercised over suppliers was not clearly defined.
Equipment Installation, Placement, Specified Requirements
The appropriate design, construction, placement, and installation of manufacturing equipment have not been ensured and equipment used in the manufacturing process does not meet specified requirements.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
Regulations FDA cited here
The 4 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 8 inspections, with how many of them cited each.
- 21 CFR 820.50Purchasing controls2 of 3
- 21 CFR 820.70Production and process controls1 of 3
- 21 CFR 820.75Process validation1 of 3
- 21 CFR 820.100Corrective and preventive action1 of 3
Other FEIs recorded as Abbott Medical
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Abbott MedicalFEI 1627487Plano, Texas, United States
Last inspected 2025-09-19NAI
Abbott MedicalFEI 3007113487Saint Paul, Minnesota, United States
Last inspected 2025-02-21NAI
Abbott MedicalFEI 2017865Sylmar, California, United States
Last inspected 2023-06-28VAI
Abbott MedicalFEI 3008377825Penang, Malaysia
Last inspected 2018-05-11NAI
Abbott MedicalFEI 2648612Caguas, Puerto Rico, United States
Last inspected 2016-02-17NAI
Abbott MedicalFEI 2938836Sunnyvale, California, United States
Last inspected 2016-01-22NAI
Abbott MedicalFEI 3009600098Westford, Massachusetts, United States
Last inspected 2014-01-29NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3006705815 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3006705815. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.