So-Low Environmental Equipment Co. Inc.
10310 Spartan Dr, Cincinnati, Ohio, United States
FEI 1526542
FDA has 3 inspections on file for So-Low Environmental Equipment Co. Inc. in Cincinnati, Ohio (FEI 1526542), from 2009 to 2014. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Apr 17, 2013. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Six regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 3, 2009–2014
- 1 OAI · 1 VAI · 1 NAI
- Inspections with 483 observations shown
- 2 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices14 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are produced in accordance with 21 CFR 820.
Show 9 more citations
Quality Audits - defined intervals
Quality audits were not performed at defined intervals to determine whether the quality system activities and results comply with quality system procedures.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed, maintained, and implemented.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Complaints
Complaint files are not adequately maintained.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated and maintained have not been established.
Documentation
Acceptance activities were not adequately documented.
Lack of or inadequate procedures for labeling
Procedures to control labeling activities have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities were not documented and implemented.
Identification procedures, Lack of or inadequate procedures
Procedures for identifying product throughout all stages of receipt, production, packaging, distribution, and installation were not established.
Management representative
No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.
Training - Lack of or inadequate procedures
Procedures for identifying training needs were not established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Show 8 more citations
Lack of or inadequate procedures for labeling
Labeling procedures were not established.
Production processes
Production processes were not complete to ensure that a device conforms to its specifications.
Lack of or inadequate procedures
The procedures for implementing corrective and preventive actions were not documented and complete.
Nonconforming product, Lack of or inadequate procedures
Procedures were not documented for the control of products that do not conform to specifications.
No Service reports
Service reports were not documented.
Document review, approval documentation
The documentation of approval of documents does not include the document approval date and the signature of the approving official.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to specifications, methods, processes, and procedures were not established.
Management review - Lack of or inadequate procedures
Procedures for management review are not established.
Regulations FDA cited here
The 16 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 2
- 21 CFR 820.20Management responsibility2 of 2
- 21 CFR 820.25Personnel2 of 2
- 21 CFR 820.70Production and process controls2 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 2
- 21 CFR 820.90Nonconforming product2 of 2
- 21 CFR 820.100Corrective and preventive action2 of 2
- 21 CFR 820.120Device labeling2 of 2
Show all 16
- 21 CFR 820.22Quality audit1 of 2
- 21 CFR 820.40[Reserved]1 of 2
- 21 CFR 820.50Purchasing controls1 of 2
- 21 CFR 820.60Identification1 of 2
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 2
- 21 CFR 820.75Process validation1 of 2
- 21 CFR 820.198Complaint files1 of 2
- 21 CFR 820.200Servicing1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1526542 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1526542. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.