Philips Medical Systems (Cleveland) Inc
595 Miner Rd, Cleveland, Ohio, United States
FEI 1525965
FDA has 10 inspections on file for Philips Medical Systems (Cleveland) Inc in Cleveland, Ohio (FEI 1525965), from 2008 to 2019. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published one citation from this inspection, under 21 CFR Part 820, the device quality system rules. It is listed below.
The most recent OAI inspection ended on Aug 18, 2017. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
12 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 10, 2008–2019
- 2 OAI · 6 VAI · 2 NAI
- Inspections with 483 observations shown
- 7 of 10
- 1 posted 483 not matched to an inspection
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 1 request, latest Jun 10, 2022
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
10 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Not approved or obsolete document retrieval
Documents that were not approved were observed at a location where they are being used.
Evaluation, timeliness, identification
Complaints representing events that are MDR reportable were not promptly reviewed, evaluated, and investigated by a designated individual.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of System for Determining MDR Events
The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate and is incomplete.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Show 4 more citations
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated and investigated where necessary.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedures for product review,disposition lack of/inadequate
Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Show 2 more citations
Design validation - software validation not performed
Validation of device software is inadequate.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices17 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.
Evaluation, timeliness, identification
Complaints representing events that are MDR reportable were not promptly reviewed, evaluated, and investigated by a designated individual.
Design review - Lack of or inadequate procedures
Procedures for design review have not been adequately established.
Personnel
Personnel do not have the necessary training to perform their jobs.
Time to report - 10 days
A report of the required information regarding device correction and removal actions was not sent to FDA within 10 days of initiating the correction or removal.
Show 12 more citations
Lack of or inadequate In-process acceptance procedures
Procedures for the control of in-process product have not been adequately established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Not approved or obsolete document retrieval
Documents that were not approved were observed at a location where they are being used.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate and is incomplete.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Supplier oversight
The type and extent of control to be exercised over the product and suppliers was not clearly defined.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Procedures not adequately established or maintained
Document control procedures have not been adequately maintained.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Procedures not adequately established or maintained
Document control procedures have not been adequately maintained.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design verification - Lack of or inadequate procedures
Design verification did not confirm that the design output meets the design input requirements.
Design validation- Lack of or inadequate procedures
Procedures for validating the device design were not implemented.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Design plans- updated
Design plans were not updated as needed as the design and development evolved.
Uniform and timely processing
Complaint handling procedures have not been defined and implemented to ensure that all complaints are processed in a uniform and timely manner.
Show 3 more citations
DMR production process specifications
The device master record does not include or refer to the location of all production and process specifications.
Documentation
Corrective and preventive action activities have not been documented, including investigations of causes of nonconformities and the actions needed to correct or prevent recurrence of nonconforming product and other quality problems.
Justification for not reporting
A justification for not reporting the correction or removal action to FDA that included follow-ups and reviews by a designated person was not included in the record.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Jun 10, 2022. It asked for a Form 483.
Inspection date not given
1 request. None of FDA’s log entries for these requests say records were released.
- Jun 10, 2022FDA-FOIA-2022-4282Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 16 sections of 21 CFR in FDA’s published citations for 7 of this facility’s 10 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action5 of 7
- 21 CFR 820.198Complaint files5 of 7
- 21 CFR 820.30Design controls4 of 7
- 21 CFR 820.40[Reserved]3 of 7
- 21 CFR 820.50Purchasing controls3 of 7
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?2 of 7
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 7
- 21 CFR 820.90Nonconforming product2 of 7
Show all 16
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 7
- 21 CFR 806.10Reports of corrections and removals1 of 7
- 21 CFR 806.20Records of corrections and removals not required to be reported1 of 7
- 21 CFR 820.20Management responsibility1 of 7
- 21 CFR 820.25Personnel1 of 7
- 21 CFR 820.70Production and process controls1 of 7
- 21 CFR 820.75Process validation1 of 7
- 21 CFR 820.181Device master record1 of 7
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Other FEIs recorded as Philips Medical Systems (Cleveland) Inc
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Philips Medical Systems (Cleveland) IncFEI 3018004899Gainesville, Florida, United States
Last inspected 2025-01-16NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1525965 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1525965. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.