Angiodynamics, Inc.
603 Queensbury Ave, Queensbury, New York, United States
FEI 1319211
FDA has 7 inspections on file for Angiodynamics, Inc. in Queensbury, New York (FEI 1319211), from 2009 to 2025. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Three FOIA requests have asked FDA for this inspection's 483.
The most recent OAI inspection ended on Sep 24, 2012. FDA has issued 2 warning letters under this FEI, the latest on (CDRH). 2 import refusals are filed under this FEI, the latest on . No court action is filed under this FEI.
Seven regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 7, 2009–2025
- 3 OAI · 2 VAI · 2 NAI
- Inspections with 483 observations shown
- 6 of 7
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest Mar 3, 2025
- Import refusals
- 2 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Records of complaint investigation
Records of complaint investigations do not include required information.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Availability
Required records were not made readily available for review and copying by the FDA.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Design input - documentation
Design input requirements were not adequately documented.
Lack of Written MDR Procedures
Written MDR procedures have not been implemented.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Show 1 more citation
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures for labeling
Procedures to control labeling activities have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Review and evaluation for investigation
Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.
Show 7 more citations
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Incorrect translation to production specifications
The device design was not correctly translated into production specifications.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Report of violation of the Act (see 803.52(e)(9))
A violation of the FD&C Act involving a device which might present a risk to health was not reported to FDA.
Lack of or inadequate procedures for distribution
Procedures for control and distribution of finished devices have not been adequately established.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design input - documentation
Design input requirements were not adequately documented.
Review and evaluation for investigation
Not all complaints have been adequately reviewed and evaluated to determine whether an investigation is necessary.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication, and evaluation of events that may be subject to medical device reporting requirements.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigation procedures
The corrective and preventive action procedures addressing the investigation of the cause of nonconformities relating to product, processes, and the quality system were not defined, complete, and implemented.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication, and evaluation of events that may be subject to medical device reporting requirements.
Uniform and timely processing
Complaint handling procedures have not been defined and implemented to ensure that all complaints are processed in a uniform and timely manner.
General
Complaint handling procedures for receiving and evaluating complaints have not been defined, completed, and implemented.
Equipment suitability & capability
Certain test equipment is not suitable for its intended purposes or capable of producing valid results.
Show 7 more citations
Environmental control Lack of or inadequate procedures
Appropriate procedures have not been defined for controlling environmental conditions.
Control/distribution procedures
Procedures for the control and distribution of finished devices have not been implemented to ensure that only devices approved for release are distributed.
Design validation- Lack of or inadequate procedures
Procedures for validating the device design were not complete.
Design output procedures
Procedures for defining and documenting design output in terms that allow an adequate evaluation of conformance to design input requirements were not documented.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
CAPA verification/validation of corrective/preventive action
Corrective and preventive actions have not been verified or validated to ensure that the action is effective and does not adversely affect the finished device.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on Mar 3, 2025. Each one asked for a Form 483.
Inspection ending Jan 10, 2025NAI
3 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Mar 3, 2025FDA-FOIA-2025-1917Form 483Closed
- Mar 3, 2025FDA-FOIA-2025-1903Form 483· 1 of 5 sitesClosed
- Feb 26, 2025FDA-FOIA-2025-1777Form 483· 1 of 2 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 16 sections of 21 CFR in FDA’s published citations for 6 of this facility’s 7 inspections, with how many of them cited each.
- 21 CFR 820.198Complaint files5 of 6
- 21 CFR 820.30Design controls4 of 6
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?3 of 6
- 21 CFR 806.10Reports of corrections and removals3 of 6
- 21 CFR 820.100Corrective and preventive action3 of 6
- 21 CFR 820.70Production and process controls2 of 6
- 21 CFR 820.160Distribution2 of 6
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 6
Show all 16
- 21 CFR 820.25Personnel1 of 6
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 6
- 21 CFR 820.75Process validation1 of 6
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 6
- 21 CFR 820.90Nonconforming product1 of 6
- 21 CFR 820.120Device labeling1 of 6
- 21 CFR 820.180General requirements1 of 6
- 21 CFR 820.184Device history record1 of 6
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1319211, one row per refused product line.
2 refusals from 2025-05-16 to 2025-12-04 · charged NOT LISTED on 1, NO 510(K) on 1 · no FDA sample analysis
FDA firm profile- 2025-12-04
CATHETER, PERCUTANEOUS, INTRAVASCULAR, LONG TERM80LJS
NOT LISTED 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of West Coast Imports799-9778664-7/10001/1
- 2025-05-16
GENERAL PURPOSE LASER PRODUCTS, NON-MEDICAL95RDW
NON STD 536(a),(b); NON STANDARD
Division of Southeast Imports799-7638247-4/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1319211 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1319211. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.