Skip to content

SunMed Holdings, LLC

2710 Northridge Dr Nw Ste A, Grand Rapids, Michigan, United States

FEI 1314417

FDA has 18 inspections on file for SunMed Holdings, LLC in Grand Rapids, Michigan (FEI 1314417), from 2008 to 2023. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 3 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.

The most recent OAI inspection ended on Mar 2, 2015. FDA issued a warning letter on (CDRH). 2 import refusals are filed under this FEI, the latest on . No court action is filed under this FEI.

Four regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
18, 2008–2023
2 OAI · 8 VAI · 8 NAI
Inspections with 483 observations shown
7 of 18
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Oct 16, 2023
Import refusals
2 refusals
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

18 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate.

    2. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  2. NAINo objectionable conditions that need follow-upDevices · Project Evaluation: DevicesNone
  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

  5. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  6. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation of validated process performance

      There is inadequate documentation of monitoring and control methods and data for a validated process.

    2. Documentation

      Process validation activities have not been adequately documented.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    4. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    5. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to a specification and process have not been adequately established.

    Show 4 more citations
    1. Acceptable supplier records, inadequate records

      Records of acceptable suppliers have not been adequately established.

    2. Lack of Written MDR Procedures

      Written MDR procedures have not been implemented.

    3. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    4. Design validation - user needs and intended uses

      Design validation did not ensure the device conforms to defined user needs and intended uses.

  7. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
  8. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  9. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
  10. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  11. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  12. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  13. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Review and evaluation for investigation

      Not all complaints have been adequately reviewed and evaluated to determine whether an investigation is necessary.

    2. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    3. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been adequately established.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    5. Sampling methods - Lack of or inadequate procedures

      Procedures to ensure sampling methods are adequate for their intended use have not been adequately established.

    Show 1 more citation
    1. Events

      A description of the events that led to the correction or removal actions was not contained in the record.

  14. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  15. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Procedures

      The procedures addressing documentation of corrective and preventive action activities were not implemented.

    2. General

      Complaint handling procedures for reviewing complaints have not been implemented.

    3. Processing MDRs {see also 803, 804}

      Complaint handling procedures have not been documented to ensure that all complaints are evaluated to determine whether the complaint should be filed as a Medical Device Report.

  16. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    2. Equipment control activity documentation

      Equipment inspections and maintenance activities have not been documented.

    3. Management review - defined interval, sufficient frequency

      Management reviews were not conducted at defined intervals.

    4. General

      Complaint handling procedures for receiving complaints have not been implemented.

    5. Identification of actions needed

      Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.

  17. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Report of death, injury

      The importer failed to submit a report to FDA on FDA form 3500A, with a copy to the manufacturer, within 30 days after receiving information that one of its marketed devices may have caused or contributed to a serious injury.

    2. Notification of extended action

      No amended notification with required information was submitted within 10 working days of initiating an extension of the correction or removal action.

    3. Evaluation of suppliers, contractors, etc., requirements

      Adequate quality requirements that must be met by suppliers were not established and documented.

    4. Results of investigation

      Records of complaint investigations do not include the results of the investigation.

  18. NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Oct 16, 2023. It asked for a Form 483.

  • Inspection ending Aug 30, 2023VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Oct 16, 2023FDA-FOIA-2023-9065Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 16 sections of 21 CFR in FDA’s published citations for 7 of this facility’s 18 inspections, with how many of them cited each.

Show all 16

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1314417, one row per refused product line.

2 refusals on 2017-12-27 · charged NOT LISTED on 2 · no FDA sample analysis

FDA firm profile
  • 2017-12-27

    DEVICE, PRESSURE APPLYING89KNM

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3808850-4/51/1

  • 2017-12-27

    DEVICE, PRESSURE APPLYING89KNM

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3808850-4/41/1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1314417 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1314417. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.