Siemens Healthcare Diagnostics, Inc.

333 Coney St, East Walpole, Massachusetts, United States

FEI 1219913

FDA has 8 inspections on file for Siemens Healthcare Diagnostics, Inc. in East Walpole, Massachusetts (FEI 1219913), from 2009 to 2022. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 2 citations from this inspection: 1 under 21 CFR Part 803, the device adverse event reporting rules, and 1 under 21 CFR Part 820, the device quality system rules. They are listed below.

The most recent OAI inspection ended on Jan 4, 2012. FDA issued a warning letter on (CDRH). 8 import refusals are filed under this FEI, the latest on . No court action is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 820.100(a), 21 CFR 803.50(a)(2) and 21 CFR 820.198(a).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
8, 2009–2022
1 OAI · 4 VAI · 3 NAI
Inspections with 483 observations shown
4 of 8
Form 483s posted by FDA
None held for this FEI
Import refusals
8 refusals
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Maintenance schedule, Lack of or inadequate schedule

      Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.

    2. Software validation for automated processes

      Software used as part of production has not been adequately validated for its intended use according to an established protocol.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities have not been adequately established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Contamination control, Lack of or inadequate procedures

      Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been adequately established.

    2. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Statistical techniques - Lack of or inadequate procedures

      Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.

    5. Complaints

      Complaint files are not adequately maintained.

  4. NAINo objectionable conditions that need follow-upDevices · Project Evaluation: DevicesNone
  5. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  6. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices13 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design verification - output does not meet input requirement

      Design verification does not confirm that design output meets design input requirements.

    2. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    Show 8 more citations
    1. Submission of Supplement to Individual Report Within One Month

      A supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.

    2. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been adequately established.

    3. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    4. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    5. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    6. Identification procedures, Lack of or inadequate procedures

      Procedures for identifying product during all stages of receipt, production, distribution, and installation have not been adequately established.

    7. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    8. Justification for not reporting

      A justification for not reporting the correction or removal action to FDA that included conclusions, follow-ups, and reviews by a designated person was not included in the record.

  7. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
  8. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone

Regulations FDA cited here

The 14 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 8 inspections, with how many of them cited each.

  • 21 CFR 820.100Corrective and preventive action4 of 4
  • 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?3 of 4
  • 21 CFR 820.70Production and process controls2 of 4
  • 21 CFR 820.198Complaint files2 of 4
  • 21 CFR 803.56If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?1 of 4
  • 21 CFR 806.10Reports of corrections and removals1 of 4
  • 21 CFR 806.20Records of corrections and removals not required to be reported1 of 4
  • 21 CFR 820.30Design controls1 of 4
Show all 14

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1219913, one row per refused product line.

8 refusals on 2018-12-04 · charged DIRECTIONS on 8, DV QUALITY on 8 · no FDA sample analysis

FDA firm profile
  • 2018-12-04

    INSTRUMENT, BIOPSY78KNW

    DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING

    DV QUALITY 501(c); 801(a)(3) Adulteration

    Division of Southeast ImportsAEK-5354644-3/91/1

  • 2018-12-04

    RADIOIMMUNOASSAY, TOTAL TRIIODOTHYRONINE75CDP

    DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING

    DV QUALITY 501(c); 801(a)(3) Adulteration

    Division of Southeast ImportsAEK-5354644-3/81/1

  • 2018-12-04

    ACID, FOLIC, RADIOIMMUNOASSAY75CGN

    DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING

    DV QUALITY 501(c); 801(a)(3) Adulteration

    Division of Southeast ImportsAEK-5354644-3/61/1

  • 2018-12-04

    TEST, NATRIURETIC PEPTIDE75NBC

    DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING

    DV QUALITY 501(c); 801(a)(3) Adulteration

    Division of Southeast ImportsAEK-5354644-3/51/1

  • 2018-12-04

    HEPATITIS A TEST (ANTIBODY AND IGM ANTIBODY)83LOL

    DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING

    DV QUALITY 501(c); 801(a)(3) Adulteration

    Division of Southeast ImportsAEK-5354644-3/131/1

Show 3 more
  • 2018-12-04

    RADIOIMMUNOASSAY, HUMAN CHORIONIC GONADOTROPIN75JHI

    DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING

    DV QUALITY 501(c); 801(a)(3) Adulteration

    Division of Southeast ImportsAEK-5354644-3/121/1

  • 2018-12-04

    TEST, HIV DETECTION82MZF

    DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING

    DV QUALITY 501(c); 801(a)(3) Adulteration

    Division of Southeast ImportsAEK-5354644-3/111/1

  • 2018-12-04

    RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)75CFT

    DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING

    DV QUALITY 501(c); 801(a)(3) Adulteration

    Division of Southeast ImportsAEK-5354644-3/101/1

Other FEIs recorded as Siemens Healthcare Diagnostics, Inc.

Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1219913 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1219913. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.