Soft Computer Consultants, Inc.

5400 Tech Data Dr, Clearwater, Florida, United States

FEI 1000307651

FDA has 4 inspections on file for Soft Computer Consultants, Inc. in Clearwater, Florida (FEI 1000307651), from 2009 to 2016. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 3 citations from this inspection: 2 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 806, the device correction and removal reporting rules. They are listed below.

The most recent OAI inspection ended on Jan 15, 2015. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Seven regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
4, 2009–2016
2 OAI · 2 VAI · 0 NAI
Inspections with 483 observations shown
4 of 4
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    3. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    2. Design input - Lack of or inadequate procedures

      Procedures for design input have not been adequately established.

    3. Events

      A description of the events that led to the correction or removal actions was not contained in the record.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    5. Lack of or inadequate organizational structure

      The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.

    Show 3 more citations
    1. Design review - Lack of or inadequate procedures

      Procedures for design review have not been adequately established.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    3. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

  3. OAISignificant objectionable conditions; FDA recommended regulatory actionBiologics · Blood and Blood Products18 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    2. Design validation - software validation not performed

      Validation of device software is incomplete.

    3. Info evaluated to determine if event was reportable

      The written MDR procedure does not include documentation and recordkeeping requirements for all information that was evaluated to determine if an event was reportable.

    4. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by suppliers and contractors have not been adequately established.

    5. Lack of or inadequate organizational structure

      The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.

    Show 13 more citations
    1. Design output - documentation

      Design output was not adequately documented before release.

    2. Design validation - Risk analysis

      Results of the design risk analysis were not adequately documented.

    3. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate.

    4. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system with sufficient frequency.

    5. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    6. Design verification - output does not meet input requirement

      Design verification does not confirm that design output meets design input requirements.

    7. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    8. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    9. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    10. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.

    11. QSR

      The quality system record has not been maintained

    12. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    13. Design input - documentation

      Design input requirements were not adequately documented.

  4. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Identification of actions needed

      Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.

    2. Analysis of data/reports from data sources

      Not all reports from quality data sources are analyzed to identify existing and potential causes of nonconforming product and other quality problems.

    3. Quality Audit/Reaudit - documentation

      The dates of quality audits have not been documented.

    4. Analysis of data sources

      Appropriate sources of quality data are not adequately analyzed to identify existing and potential causes of nonconforming product and other quality problems.

    5. CAPA verification/validation of corrective/preventive action

      Corrective and preventive actions have not been verified or validated to ensure that the action is effective and does not adversely affect the finished device.

    Show 7 more citations
    1. Identifying corrective & preventive actions

      The procedures addressing identification of corrective and preventive actions were not implemented.

    2. Management review of CAPA

      Information on identified quality problems and corrective actions has not been submitted for review by management.

    3. Info evaluated to determine if event was reportable

      The written MDR procedure does not include documentation and recordkeeping requirements for all information that was evaluated to determine if an event was reportable.

    4. Complete files maintained

      Complete complaint files are not maintained.

    5. Lack of or inadequate final acceptance procedures

      Procedures for acceptance or rejection of finished device production runs, lots, or batches were not complete.

    6. Equipment suitability & capability

      Certain test equipment is not suitable for its intended purposes or capable of producing valid results.

    7. Essential design outputs

      Design outputs that are essential for the proper functioning of the device were not identified.

Regulations FDA cited here

The 16 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.

Show all 16

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000307651 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000307651. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.