Avanos Medical, Inc.

5405 Windward Pkwy, Alpharetta, Georgia, United States

FEI 3011270181

FDA has 12 inspections on file for Avanos Medical, Inc. in Alpharetta, Georgia (FEI 3011270181), from 2009 to 2023. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 3 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Three of the five FOIA requests for this site have asked FDA for this inspection's 483.

The most recent OAI inspection ended on Jan 4, 2013. FDA has issued 2 warning letters under this FEI, the latest on (CDRH). 1 import refusal is filed under this FEI, on . No court action is filed under this FEI.

Eight regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
12, 2009–2023
3 OAI · 6 VAI · 3 NAI
Inspections with 483 observations shown
9 of 12
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
5 requests, latest FY2023
Warning letters
· CDRH ↗
2 on file
Import refusals
1 refusal
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

12 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures for monitoring and control of process parameters for a validated process have not been adequately established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. MDR not evaluated per 820.198

      A submitted MDR event was not evaluated in accordance with the requirements of the regulations regarding Complaint Files.

  4. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Device Available for Evaluation

      An individual medical device manufacturer report submitted per FDA Form 3500A did not indicate in Block D the date the device was returned to the manufacturer.

    2. Patient Gender

      An individual medical device manufacturer report submitted per FDA Form 3500A did not contain in Block A the patient's gender.

    3. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    4. Not approved or obsolete document retrieval

      Documents that were obsolete were observed at a location where they are being used.

  6. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  7. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been established to control product that does not conform to specified requirements.

    4. Documentation

      Process validation activities and results have not been adequately documented.

    5. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    Show 1 more citation
    1. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate.

  8. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Document locations, Dissemination, etc.

      Documents were not available at all locations for which they are designated, used, or otherwise necessary.

    2. Design validation - production units

      The design was not validated under defined operating conditions.

    3. Design input - documentation

      Design input requirements were not adequately documented.

    4. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been established.

    5. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.

    Show 5 more citations
    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    3. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    4. Design verification - output does not meet input requirement

      Design verification does not confirm that design output meets design input requirements.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  9. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design Validation - Risk analysis not performed/inadequate

      Risk analysis was not performed, is inadequate, and is incomplete.

    2. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    3. Personnel

      Personnel do not have the necessary training to perform their jobs.

    4. Acceptable supplier records, inadequate records

      Records of acceptable suppliers have not been adequately established.

    5. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    Show 3 more citations
    1. Review and evaluation for investigation

      Not all complaints have been adequately reviewed and evaluated to determine whether an investigation is necessary.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    3. Design validation - user needs and intended uses

      Design validation did not ensure the device conforms to defined user needs and intended uses.

  10. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation - documentation

      The results of design validation, including identification of the design, method(s), the date, and the individual(s) performing validation, were not adequately documented in the design history file.

    2. Supplier notification of changes

      There is no agreement with suppliers to notify you of changes in the product or service.

    3. Review and evaluation for investigation

      Not all complaints have been adequately reviewed and evaluated to determine whether an investigation is necessary.

    4. Complaints

      Complaint files are not adequately maintained.

  11. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Criteria for workmanship

      Criteria for workmanship have not been expressed in documented standards or by means of identified and approved representative samples.

    2. Lack of or inadequate In-process acceptance procedures

      Acceptance procedures to ensure that specified requirements for in-process product are met were not complete.

    3. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

  12. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Five requests to FDA under the Freedom of Information Act named this site, most recently on FY2023.

  • Inspection ending Dec 15, 2023VAI

    3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Aug 28, 2026. FDAgent holds no copy that FDA posted.

    1. Jun 30, 2026FDA-FOIA-2026-6208Form 483· 1 of 5 sitesRecords released
    2. Jun 11, 2026FDA-FOIA-2026-5731Form 483· 1 of 5 sitesRecords released
    3. Jan 3, 2024FDA-FOIA-2024-60Form 483Closed
  • Inspection ending Nov 14, 2018

    1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.

    1. FY2023FDA-FOIA-2023-7889Inspection recordsClosed over fees
  • Inspection date not given

    1 request. None of FDA’s log entries for these requests say records were released.

    1. Oct 6, 2020FDA-FOIA-2020-7090Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 14 sections of 21 CFR in FDA’s published citations for 9 of this facility’s 12 inspections, with how many of them cited each.

Show all 14
  • 21 CFR 820.75Process validation2 of 9
  • 21 CFR 803.18What are the requirements for establishing and maintaining MDR files or records that apply to me?1 of 9
  • 21 CFR 803.52If I am a manufacturer, what information must I submit in my individual adverse event reports?1 of 9
  • 21 CFR 820.25Personnel1 of 9
  • 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 9
  • 21 CFR 820.90Nonconforming product1 of 9

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3011270181, one row per refused product line.

1 refusal on 2022-06-14 · charged NOT LISTED on 1 · no FDA sample analysis

FDA firm profile
  • 2022-06-14

    GENERATOR, LESION, RADIOFREQUENCY84GXD

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast ImportsAEK-1274395-9/11/1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3011270181 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3011270181. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.