Respire Medical Holdings, LLC
63 Flushing Ave Unit 303 Ste 601, Brooklyn, New York, United States
FEI 3008937561
FDA has 6 inspections on file for Respire Medical Holdings, LLC in Brooklyn, New York (FEI 3008937561), from 2011 to 2022. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 6 citations from this inspection: 5 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Four regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 6, 2011–2022
- 0 OAI · 3 VAI · 3 NAI
- Inspections with 483 observations shown
- 3 of 6
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DMR - not or inadequately maintained
A device master record has not been maintained.
Training records
Personnel training is not documented.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Show 1 more citation
Evaluation and Selection, Suppliers, Contractors, etc.
Potential suppliers were not selected based on their ability to meet specified requirements.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Periodic equipment inspections
Periodic inspections of equipment were not conducted to ensure adherence to applicable maintenance schedules.
Production processes
Production processes were not conducted and controlled to ensure that a device conforms to its specifications.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Incoming acceptance records, documentation
Acceptance or rejection of incoming product was not documented.
Show 6 more citations
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Documentation
Process validation activities and results have not been adequately documented.
Design plans - Lack of or inadequate
Design plans that describe or reference the design and development activities and define responsibility for implementation have not been adequately established.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Design validation - production units
The design was not validated under defined operating conditions.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
Regulations FDA cited here
The 13 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?3 of 3
- 21 CFR 820.30Design controls3 of 3
- 21 CFR 820.198Complaint files2 of 3
- 21 CFR 806.10Reports of corrections and removals1 of 3
- 21 CFR 820.25Personnel1 of 3
- 21 CFR 820.40[Reserved]1 of 3
- 21 CFR 820.50Purchasing controls1 of 3
- 21 CFR 820.70Production and process controls1 of 3
Show all 13
- 21 CFR 820.75Process validation1 of 3
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 3
- 21 CFR 820.90Nonconforming product1 of 3
- 21 CFR 820.100Corrective and preventive action1 of 3
- 21 CFR 820.181Device master record1 of 3
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3008937561 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3008937561. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.