BPI Labs
97 S Red Willow Rd, Evanston, Wyoming, United States
FEI 3008875046DUNS 197130958
Company: Ally Labs · Linked by exact D-U-N-S match (197130958) in FDA registration data
FDA has 2 inspections on file for BPI Labs in Evanston, Wyoming (FEI 3008875046), from 2019 to 2022. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 12 citations from this inspection, all under 21 CFR Part 211, the manufacturing rules for finished drugs. They are listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, through Dec 31, 2026. Its registrant is BPI Labs, DBA Ally Labs (D-U-N-S 197130958).
Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 2, 2019–2022
- 0 OAI · 2 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Jul 16, 2024
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Records reviewed annually
Records are not maintained so that data therein can be reviewed at least annually to evaluate the quality standards of each drug product to determine the need for changes in specifications or manufacturing or control procedures.
Procedures not in writing, fully followed
The responsibilities and procedures applicable to the quality control unit are not fully followed.
Procedures for non-sterile drug products
Procedures designed to prevent objectionable microorganisms in drug products not required to be sterile are not established and written.
Representative Samples Criteria
The number of containers to be sampled is not based upon appropriate criteria.
Absence of Written Procedures
There are no written procedures for production and process controls designed to assure that the drug products have the identity, strength, quality, and purity they purport or are represented to possess.
Show 7 more citations
Identification of Equipment and Lines
The batch production and control records are deficient in that they do not include the identity of major equipment used.
Strict control not exercised over labeling issued
Strict control is not exercised over labeling issued for use in drug product labeling operations.
Training--operations, GMPs, written procedures
Employees are not given training in the particular operations they perform as part of their function and current good manufacturing practices.
Stability sample storage conditions described
The written stability program for drug products does not describe the storage conditions for samples retained for testing.
Lab controls established, including changes
The establishment of laboratory control mechanisms including any changes thereto, are not drafted by the appropriate organizational unit. Electronic records are used, but they do not meet system access limitation, audit trail, systems documentation control, signature manifestation and signature to record linking requirements to ensure that they are trustworthy, reliable and generally equivalent to paper records.
Approve or reject procedures or specs
The quality control unit lacks responsibility to approve all procedures or specifications impacting on the identity, strength, quality and purity of drug products.
Actual vs. theoretical yields not determined
Actual yield and percentages of theoretical yield are not determined at the conclusion of each appropriate phase of manufacturing of the drug product.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures for monitoring and control of process parameters for a validated process have not been established.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Show 8 more citations
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Training records
Personnel training is not documented.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
Equipment control activity documentation
Equipment maintenance activities have not been documented.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been established.
Lack of procedures, or not maintained
Document control procedures have not been established.
Management review - Lack of or inadequate procedures
Procedures for management review have not been established.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Jul 16, 2024. Each one asked for a Form 483.
Inspection ending Mar 10, 2022VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Apr 27, 2022FDA-FOIA-2022-3028Form 483Closed
Inspection date not given
1 request. None of FDA’s log entries for these requests say records were released.
- Jul 16, 2024FDA-FOIA-2024-6213Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 24 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 211.22Responsibilities of quality control unit1 of 2
- 21 CFR 211.25Personnel qualifications1 of 2
- 21 CFR 211.84Testing and approval or rejection of components, drug product containers, and closures1 of 2
- 21 CFR 211.100Written procedures; deviations1 of 2
- 21 CFR 211.103Calculation of yield1 of 2
- 21 CFR 211.113Control of microbiological contamination1 of 2
- 21 CFR 211.125Labeling issuance1 of 2
- 21 CFR 211.160General requirements1 of 2
Show all 24
- 21 CFR 211.166Stability testing1 of 2
- 21 CFR 211.180General requirements1 of 2
- 21 CFR 211.188Batch production and control records1 of 2
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
- 21 CFR 820.20Management responsibility1 of 2
- 21 CFR 820.22Quality audit1 of 2
- 21 CFR 820.25Personnel1 of 2
- 21 CFR 820.30Design controls1 of 2
- 21 CFR 820.40[Reserved]1 of 2
- 21 CFR 820.50Purchasing controls1 of 2
- 21 CFR 820.70Production and process controls1 of 2
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 2
- 21 CFR 820.75Process validation1 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.184Device history record1 of 2
FDA drug establishment registration
FDA registration search- Registered operations
- Manufacturing
- Registration runs through
- Dec 31, 2026
- Registrant
- BPI Labs, DBA Ally LabsD-U-N-S 197130958
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3008875046 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3008875046. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.