B & D Dental Corp.
2371 S Presidents Dr Ste E, West Valley City, Utah, United States
FEI 3008190898
FDA has 5 inspections on file for B & D Dental Corp. in West Valley City, Utah (FEI 3008190898), from 2013 to 2025. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 4 citations from this inspection: 3 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 807, the device registration and listing rules. They are listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Six regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 5, 2013–2025
- 0 OAI · 3 VAI · 2 NAI
- Inspections with 483 observations shown
- 3 of 5
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Feb 12, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Documentation
Process validation activities and results have not been documented.
Devices not listed
The owner or operator of a manufacturer did not submit the required device listing information for the establishment's device(s).
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices15 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documented evaluation
The evaluation of potential suppliers and contractors was not documented.
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Equipment control activity documentation
Equipment maintenance activities have not been documented.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Show 10 more citations
Lack of or inadequate procedures for storage.
Procedures for the control of storage areas and stock rooms have not been adequately established.
DMR - not or inadequately maintained
A device master record has not been maintained.
Lack of or inadequate DHR procedures
Procedures for device history records have not been established.
Training records
Personnel training is not documented.
Complaints
Complaint files are not adequately maintained.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Product rework documentation, DHR {see also 820.184}
Rework and reevaluation activities have not been fully documented in the device history record.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been established.
Incoming acceptance records, documentation
Acceptance or rejection of incoming product was not documented.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Document change records, maintained.
Records of changes to documents were not adequately maintained.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Show 5 more citations
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Remedial action
When test/measurement equipment was found to not meet accuracy and precision limits, inadequate action was taken to evaluate whether there was any adverse effect on the device's quality.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Feb 12, 2026. Each one asked for a Form 483.
Inspection ending Nov 5, 2025VAI
1 request. In the inspection history above.
- Feb 12, 2026FDA-FOIA-2026-1494Form 483· 1 of 5 sitesOpen
Inspection ending Sep 5, 2023NAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Oct 19, 2023FDA-FOIA-2023-9264Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 16 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.75Process validation3 of 3
- 21 CFR 820.100Corrective and preventive action3 of 3
- 21 CFR 820.25Personnel2 of 3
- 21 CFR 820.40[Reserved]2 of 3
- 21 CFR 820.50Purchasing controls2 of 3
- 21 CFR 820.72Inspection, measuring, and test equipment2 of 3
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 3
- 21 CFR 820.90Nonconforming product2 of 3
Show all 16
- 21 CFR 820.184Device history record2 of 3
- 21 CFR 820.198Complaint files2 of 3
- 21 CFR 820.22Quality audit1 of 3
- 21 CFR 820.30Design controls1 of 3
- 21 CFR 820.70Production and process controls1 of 3
- 21 CFR 820.150Storage1 of 3
- 21 CFR 820.181Device master record1 of 3
- 21 CFR 807.20Who must register and submit a device list?1 of 3
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3008190898 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3008190898. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.