FDA has 5 inspections on file for B & D Dental Corp. in West Valley City, Utah (FEI 3008190898), from 2013 to 2025. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 4 citations from this inspection: 3 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 807, the device registration and listing rules. They are listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Six regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
5, 2013–2025
0 OAI · 3 VAI · 2 NAI
Inspections with 483 observations shown
3 of 5
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Feb 12, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    2. Documentation

      Process validation activities and results have not been documented.

    3. Devices not listed

      The owner or operator of a manufacturer did not submit the required device listing information for the establishment's device(s).

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices15 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documented evaluation

      The evaluation of potential suppliers and contractors was not documented.

    2. Design history file

      The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.

    3. Equipment control activity documentation

      Equipment maintenance activities have not been documented.

    4. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    5. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    Show 10 more citations
    1. Lack of or inadequate procedures for storage.

      Procedures for the control of storage areas and stock rooms have not been adequately established.

    2. DMR - not or inadequately maintained

      A device master record has not been maintained.

    3. Lack of or inadequate DHR procedures

      Procedures for device history records have not been established.

    4. Training records

      Personnel training is not documented.

    5. Complaints

      Complaint files are not adequately maintained.

    6. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    7. Product rework documentation, DHR {see also 820.184}

      Rework and reevaluation activities have not been fully documented in the device history record.

    8. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been established.

    9. Incoming acceptance records, documentation

      Acceptance or rejection of incoming product was not documented.

    10. Procedures not adequately established or maintained

      Document control procedures have not been adequately established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    3. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities have not been adequately established.

    4. Document change records, maintained.

      Records of changes to documents were not adequately maintained.

    5. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    Show 5 more citations
    1. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    4. Remedial action

      When test/measurement equipment was found to not meet accuracy and precision limits, inadequate action was taken to evaluate whether there was any adverse effect on the device's quality.

    5. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

  4. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesCitations not published

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Feb 12, 2026. Each one asked for a Form 483.

  • Inspection ending Nov 5, 2025VAI

    1 request. In the inspection history above.

    1. Feb 12, 2026FDA-FOIA-2026-1494Form 483· 1 of 5 sitesOpen
  • Inspection ending Sep 5, 2023NAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Oct 19, 2023FDA-FOIA-2023-9264Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 16 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 5 inspections, with how many of them cited each.

Show all 16

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3008190898 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3008190898. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.