Skip to content

Advanced OrthoPro, Inc.

1820 N Illinois St, Indianapolis, Indiana, United States

FEI 3007912159

FDA has 4 inspections on file for Advanced OrthoPro, Inc. in Indianapolis, Indiana (FEI 3007912159), from 2009 to 2015. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

The most recent OAI inspection ended on Apr 26, 2012. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 820.198(a), 21 CFR 820.60 and 21 CFR 820.198(a)(3).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
4, 2009–2015
1 OAI · 3 VAI · 0 NAI
Inspections with 483 observations shown
4 of 4
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Periodic equipment inspection lack of or inadequate procedu

      Procedures for conducting periodic inspections to ensure adherence to equipment maintenance schedules have not been established.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of System for Timely Submission of Reports

      The written MDR procedure does not include an internal system which provides for timely transmission of complete medical device reports to FDA.

    2. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been established.

    3. Lack of procedures, or not maintained

      Document control procedures have not been maintained.

    4. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    5. Identification procedures, Lack of or inadequate procedures

      Procedures for identifying product during all stages of receipt, production, distribution, and installation have not been established.

    Show 6 more citations
    1. Lack of or inadequate DHR procedures

      Procedures for device history records have not been established.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Training records

      Personnel training is not documented.

    4. DMR - not or inadequately maintained

      A device master record has not been maintained.

    5. Management review - Lack of or inadequate procedures

      Procedures for management review have not been adequately established.

    6. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Complete files maintained

      Complete complaint files are not maintained.

    2. Availability

      Quality Systems records were not made readily available to FDA employees.

    3. Processing MDRs {see also 803, 804}

      Complaint handling procedures have not been established and defined to ensure that all complaints are evaluated to determine whether the complaint should be filed as a Medical Device Report.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design control - no procedures

      Procedures to control the design process of the device were not established, defined, documented, and implemented.

    2. Lack of or inadequate final acceptance procedures

      Procedures for acceptance or rejection of finished device production runs, lots, or batches were not established, defined, and documented.

    3. Lack of or inadequate procedures for labeling

      Labeling procedures were not established, defined, and documented.

    4. Design history file

      The design history file was not established or maintained.

    5. Lack of Written MDR Procedures

      Written MDR procedures have not been developed and maintained.

    Show 6 more citations
    1. Quality plan

      A quality plan was not complete.

    2. Procedures for product review,disposition lack of/inadequate

      Procedures that describe the review and disposition process for nonconforming product were not established, defined, and documented.

    3. Document locations, Dissemination, etc.

      Documents were not available where they were necessary.

    4. Procedures

      The procedures addressing documentation of corrective and preventive action activities were not established, defined, and documented.

    5. Identification procedures, Lack of or inadequate procedures

      Procedures for identifying product throughout all stages of receipt, production, packaging, distribution, and installation were not established, defined, and documented.

    6. Processing MDRs {see also 803, 804}

      Complaint handling procedures have not been established, defined, and documented to ensure that all complaints are evaluated to determine whether the complaint should be filed as a Medical Device Report.

Regulations FDA cited here

The 16 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.

Show all 16

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3007912159 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3007912159. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.