Compass Richmar LLC
4120 S Creek Rd, Chattanooga, Tennessee, United States
FEI 3004750362
FDA has 5 inspections on file for Compass Richmar LLC in Chattanooga, Tennessee (FEI 3004750362), from 2009 to 2016. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Jun 28, 2012. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Seven regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 5, 2009–2016
- 2 OAI · 2 VAI · 1 NAI
- Inspections with 483 observations shown
- 4 of 5
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation of software validation
Software validation activities and results for computers or automated data processing systems used as part of the quality system have not been documented.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Failure to submit
You did not submit an annual report for products requiring one.
Show 2 more citations
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.
Acceptable supplier records, inadequate records
Records of acceptable suppliers, contractors, and consultants have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Radiation Control and Health Safety Act6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers have not been adequately established.
Failure to submit, distinct marking
You did not submit a required product report prior to the introduction of the product into commerce.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Failure to submit
You did not submit an annual report for products requiring one.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Show 1 more citation
Product rework documentation, DHR {see also 820.184}
Rework and reevaluation activities have not been fully documented in the device history record.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices16 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Review and evaluation for investigation
Not all complaints have been adequately reviewed and evaluated to determine whether an investigation is necessary.
Quality audits - auditor independence
Individuals who conduct quality audits have direct responsibility for the matters being audited.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers have not been adequately established.
Lack of Written MDR Procedures
Written MDR procedures have not been implemented.
Show 11 more citations
Documentation
Acceptance activities were not adequately documented.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been adequately established.
DMR - not or inadequately maintained
A device master record has not been maintained.
Design validation - software validation not performed
Validation of device software was not performed.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Personnel
Personnel do not have the necessary training to perform their jobs.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Records not kept
There is no record maintained of a correction or removal action that was not required to be reported to FDA.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices15 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product were not defined, documented, and implemented.
Product rework documentation, DHR {see also 820.184}
Rework and reevaluation activities have not been fully documented in the device history record.
Documentation of disposition, justification, signature
There is incomplete documentation of the disposition of nonconforming product.
ID label, labeling
The device history record does not include the primary identification label and labeling for each device.
Quality Audits - defined intervals
Quality audits did not verify that the quality system is effective in fulfilling your quality system objectives.
Show 10 more citations
Document locations, Dissemination, etc.
Documents were not available where they were designated to be kept.
Incoming inspection, testing, verification
Incoming product was not adequately inspected or tested to verify conformance to specifications.
Investigation
Certain indicators of nonconformities are not investigated to determine the cause of the nonconformity.
Supplier oversight
The type and extent of control to be exercised over suppliers was not clearly defined.
Management representative
No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.
Analysis of data sources
Appropriate sources of quality data are not adequately analyzed to identify existing and potential causes of nonconforming product and other quality problems.
Lack of or inadequate procedures for handling
Procedures have not been implemented to ensure that mix-ups, damage, or other adverse effects to product do not occur during handling.
Training records
Employee training is not fully documented.
Management responsibility
Management with executive responsibility has not ensured that an adequate and effective quality system has been fully implemented and maintained at all levels of the organization.
Design changes - Lack of or Inadequate Procedures
Procedures were not followed for the of design changes before their implementation.
Regulations FDA cited here
The 18 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.50Purchasing controls4 of 4
- 21 CFR 820.100Corrective and preventive action4 of 4
- 21 CFR 820.30Design controls3 of 4
- 21 CFR 820.90Nonconforming product3 of 4
- 21 CFR 820.198Complaint files3 of 4
- 21 CFR 820.20Management responsibility2 of 4
- 21 CFR 820.22Quality audit2 of 4
- 21 CFR 820.25Personnel2 of 4
Show all 18
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 4
- 21 CFR 820.184Device history record2 of 4
- 21 CFR 1002.13Annual reports2 of 4
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 4
- 21 CFR 806.20Records of corrections and removals not required to be reported1 of 4
- 21 CFR 820.40[Reserved]1 of 4
- 21 CFR 820.70Production and process controls1 of 4
- 21 CFR 820.140Handling1 of 4
- 21 CFR 820.181Device master record1 of 4
- 21 CFR 1002.10Product reports1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004750362 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004750362. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.