Endotec, Inc.
14525 Valley View Ave Ste H, Santa Fe Springs, California, United States
FEI 3004553437
FDA has 5 inspections on file for Endotec, Inc. in Santa Fe Springs, California (FEI 3004553437), from 2009 to 2017. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
FDA issued a warning letter on (CDRH). FDA's compliance record also lists one other action under this FEI: Injunction, . No import refusal is filed under this FEI.
Nine regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 5, 2009–2017
- 3 OAI · 2 VAI · 0 NAI
- Inspections with 483 observations shown
- 5 of 5
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by contractors and consultants have not been established.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system with sufficient frequency.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Quality Audits - defined intervals
Quality audits were not performed at sufficient frequency to determine whether the quality system activities and results comply with quality system procedures.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Sampling methods - Lack of or inadequate procedures
Procedures to ensure sampling methods are adequate for their intended use have not been adequately established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Show 2 more citations
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Process validation results have not been adequately documented.
Design validation - documentation
The results of design validation, including method(s), were not documented in the design history file.
Quality Audits - defined intervals
Quality audits were not performed at defined intervals to determine whether the quality system activities and results comply with quality system procedures.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Show 8 more citations
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Procedures for product review,disposition lack of/inadequate
Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Design input - Lack of or inadequate procedures
Procedures for design input have not been adequately established.
Lack of or inadequate procedures for storage.
Procedures for the control of storage areas and stock rooms have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Management review - Lack of or inadequate procedures
Procedures for management review have not been established.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Investigation
Certain indicators of nonconformities are not investigated to determine the cause of the nonconformity.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Criteria for workmanship
Criteria for workmanship have not been expressed in documented standards or by means of identified and approved representative samples.
Not approved or obsolete document retrieval
Documents that were obsolete were observed at a location where they could be used.
Show 8 more citations
Quality audits - Lack of or inadequate procedures
Procedures for conducting quality audits were not defined and complete.
Process control procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications were not defined and implemented.
Design control - no procedures
Procedures to control the design process of the device were not implemented.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated and approved according to established procedures.
Production processes
Production processes were not controlled and monitored to ensure that a device conforms to its specifications.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not defined and implemented.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Lack of or inadequate procedures
The procedures for implementing corrective and preventive actions were not defined, complete, and implemented.
Regulations FDA cited here
The 15 sections of 21 CFR in FDA’s published citations for this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.22Quality audit4 of 5
- 21 CFR 820.50Purchasing controls4 of 5
- 21 CFR 820.30Design controls3 of 5
- 21 CFR 820.70Production and process controls3 of 5
- 21 CFR 820.75Process validation3 of 5
- 21 CFR 820.90Nonconforming product3 of 5
- 21 CFR 820.100Corrective and preventive action3 of 5
- 21 CFR 820.20Management responsibility2 of 5
Show all 15
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 5
- 21 CFR 806.10Reports of corrections and removals1 of 5
- 21 CFR 820.40[Reserved]1 of 5
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 5
- 21 CFR 820.150Storage1 of 5
- 21 CFR 820.198Complaint files1 of 5
- 21 CFR 820.250Statistical techniques1 of 5
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- InjunctionCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004553437 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004553437. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.