Alcor Scientific, Inc.
20 Thurber Blvd, Smithfield, Rhode Island, United States
FEI 3003855297
FDA has 4 inspections on file for Alcor Scientific, Inc. in Smithfield, Rhode Island (FEI 3003855297), from 2009 to 2018. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 3 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
The most recent OAI inspection ended on Dec 31, 2009. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.198(a), 21 CFR 820.30(g) and 21 CFR 820.100(b).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2009–2018
- 1 OAI · 3 VAI · 0 NAI
- Inspections with 483 observations shown
- 4 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Equipment suitability & capability
Certain inspection, measuring and test equipment is not suitable for its intended purposes.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Complaints
Complaint files are not adequately maintained.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Design validation - production units
The design was not validated under defined operating conditions.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Acceptance activities were not adequately documented.
Lack of procedures, or not maintained
Document control procedures have not been established and maintained.
Records of complaint investigation
Records of complaint investigations do not include required information.
Lack of or inadequate procedures for storage.
Procedures for the control of storage areas and stock rooms have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices14 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality audit corrective action, reaudits {see also 820.100}
Quality reaudits are not adequately performed.
Training
Employees have not been adequately trained.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Statistical techniques - Lack of or inadequate procedures
Procedures to identify valid statistical techniques needed for certain processes and certain product analyses have not been established.
DHR content
The device history record does not demonstrate the device is manufactured in accordance with the device master record.
Show 9 more citations
Documentation
Corrective and preventive action activities have not been documented, including investigations of causes of nonconformities, the verification or validation of corrective actions, and implementation of corrective and preventive actions.
Lack of or inadequate procedures for handling
Procedures have not been implemented to ensure that mix-ups, damage, or other adverse effects to product do not occur during handling.
Production processes
Production processes were not complete to ensure that a device conforms to its specifications.
Review and evaluation for investigation
Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.
Lack of or inadequate procedures for labeling
Labeling procedures were not established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not implemented.
Uniform and timely processing
Complaint handling procedures have not been established to ensure that all complaints are processed in a uniform and timely manner.
Acceptable supplier records
Records of acceptable suppliers were not maintained.
Oral complaints
Complaint handling procedures have not been established to ensure that all oral complaints are documented upon receipt.
Regulations FDA cited here
The 16 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.198Complaint files4 of 4
- 21 CFR 820.30Design controls2 of 4
- 21 CFR 820.100Corrective and preventive action2 of 4
- 21 CFR 820.22Quality audit1 of 4
- 21 CFR 820.25Personnel1 of 4
- 21 CFR 820.40[Reserved]1 of 4
- 21 CFR 820.50Purchasing controls1 of 4
- 21 CFR 820.70Production and process controls1 of 4
Show all 16
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 4
- 21 CFR 820.90Nonconforming product1 of 4
- 21 CFR 820.120Device labeling1 of 4
- 21 CFR 820.140Handling1 of 4
- 21 CFR 820.150Storage1 of 4
- 21 CFR 820.184Device history record1 of 4
- 21 CFR 820.250Statistical techniques1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3003855297 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3003855297. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.