Skip to content

Medical Measurement Systems B.V.

Colosseum 25, Enschede, Netherlands

FEI 3002957361

FDA has 2 inspections on file for Medical Measurement Systems B.V. in Enschede, Netherlands (FEI 3002957361), from 2012 to 2015. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 13 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Five regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
2, 2012–2015
1 OAI · 0 VAI · 1 NAI
Inspections with 483 observations shown
2 of 2
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices13 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Production processes

      Production processes were not developed, conducted, controlled, and monitored to ensure that a device conforms to its specifications.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Servicing - Lack of or inadequate procedures

      Procedures or instructions for performing servicing activities and verifying that servicing meets specified requirements have not been adequately established.

    4. Design verification - documentation

      The design verification results, including method(s), were not adequately documented in the design history file.

    5. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    Show 8 more citations
    1. Evaluation and Selection, Suppliers, Contractors, etc.

      Potential suppliers were not evaluated and selected based on their ability to meet specified requirements.

    2. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    3. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated, inspected, checked, and maintained have not been adequately established.

    4. Records of complaint investigation

      Records of complaint investigations do not include required information.

    5. Equipment Installation, Placement, Specified Requirements

      The appropriate design, construction, placement, and installation of manufacturing equipment have not been ensured.

    6. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been established.

    7. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    8. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DMR - not or inadequately maintained

      A device master record has not been maintained.

    2. Investigation of device failures

      Complaints involving the possible failure of a device and packaging to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.

    3. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    4. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been established.

    5. Documentation

      Corrective and preventive action activities and/or results have not been documented.

    Show 5 more citations
    1. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.

    2. Document review, approval by designated individual

      Documents were not reviewed and not approved by designated individual(s) prior to issuance .

    3. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    4. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

Regulations FDA cited here

The 14 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.

Show all 14

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3002957361 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3002957361. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.