Signal Medical Corporation
400 Pyramid Dr, Marysville, Michigan, United States
FEI 3002477077
FDA has 10 inspections on file for Signal Medical Corporation in Marysville, Michigan (FEI 3002477077), from 2009 to 2023. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
The most recent OAI inspection ended on Oct 12, 2010. FDA has issued 2 warning letters under this FEI, the latest on (CDRH). No court action or import refusal is filed under this FEI.
Four regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 10, 2009–2023
- 2 OAI · 3 VAI · 5 NAI
- Inspections with 483 observations shown
- 5 of 10
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Dec 22, 2023
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
10 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Statistical techniques - Lack of or inadequate procedures
Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Supplier notification of changes
There is no agreement with suppliers, contractors and consultants to notify you of changes in the product or service.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Show 2 more citations
Investigation of device failures
Complaints involving the possible failure of a device, labeling and packaging to meet any of its specifications were not reviewed, evaluated and investigated where necessary.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with 21 CFR 820.
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Show 7 more citations
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Software validation for automated processes
Software used as part of the quality system has not been adequately validated for its intended use according to an established protocol.
Lack of System for Determining MDR Events
The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.
Remedial action
When test/measurement equipment was found to not meet accuracy and precision limits, no action was taken to evaluate whether there was any adverse effect on the device's quality.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Quality audits - auditor independence
Individuals who conduct quality audits have direct responsibility for the matters being audited.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Process control procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications were not established.
Acceptance records
The device history record does not include complete acceptance records that demonstrate the device is manufactured in accordance with the device master record.
Statistical techniques - Lack of or inadequate procedures
Procedures to identify valid statistical techniques needed for certain processes have not been established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated and approved according to established procedures.
Process changes - review, evaluation and revalidation
A validated process was not reviewed and evaluated and revalidated when changes or process deviations occurred.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Dec 22, 2023. It asked for a Form 483.
Inspection ending Apr 10, 2023VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Dec 22, 2023FDA-FOIA-2023-11439Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 16 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 10 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls3 of 5
- 21 CFR 820.75Process validation3 of 5
- 21 CFR 820.100Corrective and preventive action3 of 5
- 21 CFR 820.50Purchasing controls2 of 5
- 21 CFR 820.70Production and process controls2 of 5
- 21 CFR 820.184Device history record2 of 5
- 21 CFR 820.250Statistical techniques2 of 5
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 5
Show all 16
- 21 CFR 806.10Reports of corrections and removals1 of 5
- 21 CFR 820.22Quality audit1 of 5
- 21 CFR 820.25Personnel1 of 5
- 21 CFR 820.40[Reserved]1 of 5
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 5
- 21 CFR 820.90Nonconforming product1 of 5
- 21 CFR 820.181Device master record1 of 5
- 21 CFR 820.198Complaint files1 of 5
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3002477077 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3002477077. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.