Skip to content

Custom Assemblies, Inc.

330 E Main St, Pine Level, North Carolina, United States

FEI 3000718854

FDA has 6 inspections on file for Custom Assemblies, Inc. in Pine Level, North Carolina (FEI 3000718854), from 2011 to 2025. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

The most recent OAI inspection ended on Jul 31, 2017. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Four regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
6, 2011–2025
2 OAI · 1 VAI · 3 NAI
Inspections with 483 observations shown
4 of 6
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Procedures not adequately established or maintained

      Document control procedures have not been adequately maintained.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    2. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    3. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to a specification have not been adequately established.

    2. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been established to control product that does not conform to specified requirements.

    3. Info evaluated to determine if event was reportable

      The written MDR procedure does not include documentation and recordkeeping requirements for all information that was evaluated to determine if an event was reportable.

    4. Personnel

      Personnel do not have the necessary education, background, training and experience to perform their jobs.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    Show 7 more citations
    1. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    2. Product rework documentation, DHR {see also 820.184}

      Rework and reevaluation activities have not been documented in the device history record.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Design control - no procedures

      Procedures for design control have not been established.

    5. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been established.

    6. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures for monitoring and control of process parameters for a validated process have not been adequately established.

    7. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

  5. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  6. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Training records

      Personnel training is not documented.

    2. Personnel

      Personnel do not have the necessary education, background, training, and experience to perform their jobs.

    3. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    4. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been adequately established.

    5. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    Show 7 more citations
    1. Lack of Written MDR Procedures

      Written MDR procedures have not been maintained and implemented.

    2. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures for monitoring and control of process parameters for a validated process have not been adequately established.

    3. Management review - Lack of or inadequate procedures

      Procedures for management review have not been adequately established.

    4. Lack of or inadequate procedures for handling

      Procedures for product handling have not been established.

    5. Process changes - review, evaluation and revalidation

      A validated process was not reviewed and evaluated when changes or process deviations occurred.

    6. Remedial action

      When test/measurement equipment was found to not meet accuracy and precision limits, no action was taken to evaluate whether there was any adverse effect on the device's quality.

    7. Documentation of validated process performance

      There is inadequate documentation of monitoring and control methods and data for a validated process.

Regulations FDA cited here

The 15 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 6 inspections, with how many of them cited each.

Show all 15

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3000718854 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3000718854. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.